K Nishizawa, H Maekoshi, Y Kamiya, Y Kobayashi, K Ohara, S Sakuma.
Abstract
A general method to align a diagnostic x-ray machine for x-ray spectrum measurement purpose was theoretically and experimentally investigated by means of the optical alignment of focal pinhole images. Focal pinhole images were obtained by using a multi-pinholed lead plate. the vertical plane, including the central axis and tube axis, was decided upon by observing the symmetry of focal images. the central axis was designated as a line through the center of focus parallel to the target surface lying in the vertical plane. A method to determine the manipulation of the central axis in any direction is presented.Entities:
Mesh:
Year: 1982 PMID: 7087916 DOI: 10.1118/1.595086