Literature DB >> 7087916

Alignment of x-ray tube focal spots for spectral measurement.

K Nishizawa, H Maekoshi, Y Kamiya, Y Kobayashi, K Ohara, S Sakuma.   

Abstract

A general method to align a diagnostic x-ray machine for x-ray spectrum measurement purpose was theoretically and experimentally investigated by means of the optical alignment of focal pinhole images. Focal pinhole images were obtained by using a multi-pinholed lead plate. the vertical plane, including the central axis and tube axis, was decided upon by observing the symmetry of focal images. the central axis was designated as a line through the center of focus parallel to the target surface lying in the vertical plane. A method to determine the manipulation of the central axis in any direction is presented.

Entities:  

Mesh:

Year:  1982        PMID: 7087916     DOI: 10.1118/1.595086

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  1 in total

1.  A convenient alignment approach for x-ray imaging experiments based on laser positioning devices.

Authors:  Zhang Da; Molly Donovan; Xizeng Wu; Hong Liu
Journal:  Med Phys       Date:  2008-11       Impact factor: 4.071

  1 in total