Literature DB >> 6887186

Femoral neck fracture fixation: rigidity of five techniques compared.

A C Mackechnie-Jarvis.   

Abstract

Artificial cadaveric femoral neck fractures were internally fixed with five different devices and subjected to cyclical loading of 0-1.0 kilonewtons (approximately one body weight) whilst in an anatomical position. Displacement of the proximal fragment was detected by a transducer and charted. Bone strength was assessed by a preliminary control loading phase on the intact bone. Efficiency of each fracture fixator could then be directly compared by the relative movement in each case. Five specimens each were tested with Moore's Pins, Trifin Nail, Garden Screws and a sliding screw-plate (OEC Ltd). By the criteria of the experiment, which put a severe shearing load on the implant, none of these devices reliably bore the representative body weight. An extended barrel-plate, which supported the sliding screw almost up to the fracture line, was then made. This device, employing some of Charnley's concepts, tolerated body weight in four cases out of five.

Entities:  

Mesh:

Year:  1983        PMID: 6887186      PMCID: PMC1439299          DOI: 10.1177/014107688307600805

Source DB:  PubMed          Journal:  J R Soc Med        ISSN: 0141-0768            Impact factor:   18.000


  8 in total

1.  STABILITY AND UNION IN SUBCAPITAL FRACTURES OF THE FEMUR.

Authors:  R S GARDEN
Journal:  J Bone Joint Surg Br       Date:  1964-11

2.  The treatment of displaced fractures of the neck of the femur by compression.

Authors:  J CHARNLEY; N J BLOCKEY; D W PURSER
Journal:  J Bone Joint Surg Br       Date:  1957-02

3.  A new principle in the operative treatment of trochanteric fractures of the femur.

Authors:  W SCHUMPELICK; P M JANTZEN
Journal:  J Bone Joint Surg Am       Date:  1955-07       Impact factor: 5.284

4.  Internal fixation versus hemiarthroplasty for the displaced subcapital fracture of the femur. A prospective randomised study.

Authors:  J M Sikorski; R Barrington
Journal:  J Bone Joint Surg Br       Date:  1981

5.  Biomechanical analysis of the sliding characteristics of compression hip screws.

Authors:  R F Kyle; T M Wright; A H Burstein
Journal:  J Bone Joint Surg Am       Date:  1980-12       Impact factor: 5.284

6.  Perosseous venography in the diagnosis of viability in subcapital fractures of the femur.

Authors:  R E Outerbridge
Journal:  Clin Orthop Relat Res       Date:  1978 Nov-Dec       Impact factor: 4.176

7.  Femoral neck fractures in young adults.

Authors:  H Kofoed
Journal:  Injury       Date:  1982-09       Impact factor: 2.586

8.  Treatment of nondisplaced fractures of the femoral neck.

Authors:  G Bentley
Journal:  Clin Orthop Relat Res       Date:  1980-10       Impact factor: 4.176

  8 in total