R R Little, J D England, H M Wiedmeyer, D E Goldstein.
Abstract
After storage of whole blood at either 4 or 20 degrees C, results for glycosylated hemoglobin by ion-exchange chromatography ("high-performance" liquid and mini-column chromatography), thiobarbituric acid colorimetry, and affinity chromatography were compared. At 4 degrees C, all methods gave acceptable results for samples stored for as long as a week. At 20 degrees C, the colorimetric and affinity methods also showed sample stability for a week or more. The ion-exchange methods were associated with a marked increase in values for glycosylated hemoglobin after a few days of storage. Evidently, care in details of sample collection and handling is especially important for ion-exchange methods, and the colorimetric and affinity methods have advantages over ion exchange in situations where long delays between sample collection and assay are unavoidable.Entities: