Application of reflectance spectroscopy to the estimation of uric acid, urea and glucose: an evaluation of the Ames Seralyzer.

J F Stevens, R G Newall.   

Abstract

An original approach to the measurement of analytes in clinical chemistry has now become available, in which dry reagent strip technology is linked to measurement by reflectance spectroscopy. The present studies have evaluated the performance of the first of these test systems-for uric acid, urea and glucose, in serum-by comparison with conventional liquid chemistry methods. Satisfactory performance in terms of both precision and accuracy was obtained for all three test systems, the current "state-of-the-art" performance criteria being met; the Seralyzer system proved reliable and easy to use in the hands of trained operators. It should find a place as a "Stat" analyser in the laboratory, in specified wards and in Health Centres.

Entities:  

Mesh:

Substances:

Year:  1983        PMID: 6822683      PMCID: PMC498096          DOI: 10.1136/jcp.36.1.9

Source DB:  PubMed          Journal:  J Clin Pathol        ISSN: 0021-9746            Impact factor:   3.411


  1 in total

1.  Analytic clinical laboratory precision--state of the art for thirty-one analysis.

Authors:  J W Ross; M D Fraser; T D Moore
Journal:  Am J Clin Pathol       Date:  1980-10       Impact factor: 2.493

  1 in total
  2 in total

1.  Measurement of the enzymes lactate dehydrogenase and creatine kinase using reflectance spectroscopy and reagent strips.

Authors:  J F Stevens; W Tsang; R G Newall
Journal:  J Clin Pathol       Date:  1983-12       Impact factor: 3.411

2.  Measurement of bilirubin, cholesterol and creatinine in serum and plasma, by solid-phase reflectance spectroscopy.

Authors:  J F Stevens; W Tsang; R G Newall
Journal:  J Clin Pathol       Date:  1983-05       Impact factor: 3.411

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.