[Value and limitations of the computed x-ray tomographic measurement of vertebral mineralization].

A M Laval-Jeantet, M Laval-Jeantet, B Roger, C Scott, P F Delmas.   

Abstract

An experimental study of the reproducibility and exactitude of densitometric measurements obtained by CT scanning showed: 1) the existence of marked variations due to factors related to the apparatus: beam energy, filtration, electronics, direction of rotation, reconstruction software; 2) that the structure of the object was also an influencing factor: variations in tissues surrounding the vertebra, and variations in bone density as a function of the plane in the vertebral body, and in fat content in the spongy tissue. Any measurement by CT scan of the mineral density of the vertebral body requires the use of a calibration phantom, strict definition of the volume to be measured, and routine correction for hypodensity induced by fat when necessary. Results under these conditions are reproducible and useful for the diagnosis of rarefying osteopathies.

Mesh:

Substances:

Year:  1984        PMID: 6716345

Source DB:  PubMed          Journal:  J Radiol        ISSN: 0221-0363


  4 in total

1.  CT image analysis of the vertebral trabecular network in vivo.

Authors:  F Chevalier; A M Laval-Jeantet; M Laval-Jeantet; C Bergot
Journal:  Calcif Tissue Int       Date:  1992-07       Impact factor: 4.333

2.  Accuracy of vertebral mineral determination by dual-energy quantitative computed tomography.

Authors:  W D Reinbold; C P Adler; W A Kalender; R Lente
Journal:  Skeletal Radiol       Date:  1991       Impact factor: 2.199

Review 3.  Noninvasive methods of bone-mass measurement.

Authors:  J P Sabatier; G Guaydier-Souquieres
Journal:  Clin Rheumatol       Date:  1989-06       Impact factor: 2.980

4.  Measurement of anisotropic vertebral trabecular bone loss during aging by quantitative image analysis.

Authors:  C Bergot; A M Laval-Jeantet; F Prêteux; A Meunier
Journal:  Calcif Tissue Int       Date:  1988-09       Impact factor: 4.333