Vaporization of mixed anesthetic liquids.

D L Bruce, H W Linde.   

Abstract

The results of erroneous filling of agent-specific anesthetic vaporizers were studied. The fraction of gas flow through the vaporizer was calculated for three vaporizers set to deliver essentially equipotent final concentrations: halothane, 1% (1.25 MAC); enflurane, 2% (1.19 MAC); and isoflurane, 1.5% (1.30 MAC). These fractional flows, at 22 degrees C, were 0.0188 for 1% halothane, 0.0615 for 2% enflurane, and 0.0295 for 1.5% isoflurane. Concentrations were calculated for cases of total filling of a vaporizer with one of the other two agents. In terms of potency of delivered agent, fourfold underdoses or overdoses could result from such errors. Refilling a 25% full vaporizer with the wrong agent then was considered. In order to calculate the concentrations of each agent that would be delivered in such a case, vapor pressures of each were determined in mixtures of two agents. Enflurane and isoflurane could not be separated satisfactorily by gas chromatography. Halothane, when mixed with enflurane or isoflurane, enhanced vaporization of each agent, as well as being somewhat more easily vaporized itself. Halothane, enflurane, and isoflurane do not form ideal solutions when mixed and the resultant vapor concentrations of each of two agents when mixed may be far from those predicted by an assumption of ideality.

Entities:  

Mesh:

Substances:

Year:  1984        PMID: 6703388     DOI: 10.1097/00000542-198404000-00010

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  5 in total

1.  Recognition of mixed anesthetic agents by mass spectrometer during anesthesia.

Authors:  C Munshi; S Dhamee; A Bardeen-Henschel; S Dhruva
Journal:  J Clin Monit       Date:  1986-04

2.  Evaluation of three transportable multigas anesthetic monitors: the Bruel & Kjaer Anesthetic Gas Monitor 1304, the Datex Capnomac Ultima, and the Nellcor N-2500.

Authors:  J Nielsen; T Kann; J T Moller
Journal:  J Clin Monit       Date:  1993-04

3.  Observations on use of wrong agent in an anesthesia agent vaporizer.

Authors:  F E Block; G T Schulte
Journal:  J Clin Monit Comput       Date:  1999-01       Impact factor: 2.502

4.  Performance of erroneously filled sevoflurane, enflurane and other agent-specific vaporizers.

Authors:  M Abel; J B Eisenkraft
Journal:  J Clin Monit       Date:  1996-03

Review 5.  Safety features in anaesthesia machine.

Authors:  M Subrahmanyam; S Mohan
Journal:  Indian J Anaesth       Date:  2013-09
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.