Literature DB >> 6703298

Evaluation of Spirolog I volume meter.

M K Chakrabarti, L Loh.   

Abstract

A volume meter (Dräger Spirolog I) is evaluated for clinical use. It works on the principle of a constant temperature anemometer and displays tidal volume, minute volume and frequency. It was found that the volume recorded was hardly affected by oxygen concentration, humidity or temperature but was affected by high concentrations of nitrous oxide and during the operation of diathermy.

Entities:  

Mesh:

Substances:

Year:  1984        PMID: 6703298     DOI: 10.1111/j.1365-2044.1984.tb07241.x

Source DB:  PubMed          Journal:  Anaesthesia        ISSN: 0003-2409            Impact factor:   6.955


  1 in total

1.  Evaluation of two oxygen analyzers by computerized data acquisition and processing.

Authors:  K Erdmann; J P Jantzen; C Etz; W F Dick
Journal:  J Clin Monit       Date:  1986-04
  1 in total