An improved microindentation technique to measure changes in properties of arterial intima during atherogenesis.

B S Gow, W D Castle, M J Legg.   

Abstract

A fully automated instrument was developed that measured the amount and time-course of indentation of arterial intima by a spherical tipped probe 25 micrometers in diameter loaded routinely by forces of 10 microN. The relationship of depth of indentation to time was non-exponential but reached an apparent asymptotic value in about 6 s. Routine indentations were recorded 10 s after loading, the relationship of indentation and force being essentially linear between 0-100 microN. The instrument is an order of magnitude more sensitive than one previously described (Gow and Vaishnav, 1975, J. Appl. Physiol. 38, 344-350) and produces indentations of 1-4 micrometers in the intima of rabbit thoracic aorta with a reproducibility of 1-2%. Preliminary usage of the new microindentor has shown increases in the compliance of the thoracic aorta intima in rabbits during cholesterol feeding.

Entities:  

Mesh:

Year:  1983        PMID: 6619161     DOI: 10.1016/0021-9290(83)90077-5

Source DB:  PubMed          Journal:  J Biomech        ISSN: 0021-9290            Impact factor:   2.712