Literature DB >> 6548989

Primary and leakage radiation calculations at 6, 10 and 25 MeV.

W R Nelson, P D LaRiviere.   

Abstract

Primary and leakage x-ray spectra for typical clinical accelerators have been calculated using the EGS Monte Carlo code for 3 energies (6, 10 and 25 MeV) and 4 angles (0 degree, 45 degrees, 90 degrees and 135 degrees). Broad-beam transmission curves have been calculated for ordinary concrete using the MORSE program with the EGS spectra as input. A simple analytical model, which is shown to agree rather well with both experimental data and with the MORSE results, is presented and initial and subsequent tenth-value layers are extracted. Finally, the photon spectrum after the concrete shield is obtained with MORSE.

Mesh:

Year:  1984        PMID: 6548989     DOI: 10.1097/00004032-198412000-00001

Source DB:  PubMed          Journal:  Health Phys        ISSN: 0017-9078            Impact factor:   1.316


  1 in total

1.  Uncertainties in linac primary barrier transmission values.

Authors:  Patrick N McDermott; Michael D Sigler; Ian P Lake; Danielle Lack
Journal:  J Appl Clin Med Phys       Date:  2022-03-02       Impact factor: 2.102

  1 in total