A T Cohen, P C Beatty, B Kay, T E Healy.
Abstract
A new method for oxygen uptake (VO2) measurement based on a constant volume closed circuit is discussed. The accuracy of the method tested in the laboratory was demonstrated to be +/- 1.8% and estimated to be +/- 2.01% in clinical use during abdominal surgery. The 0-90% rise time of the measurement to step changes in VO2 was found to be 42 s, for a change of 15 ml min-1 with a lag time of less than 30 s.Entities:
Mesh:
Substances:
Year: 1984 PMID: 6536508
Source DB: PubMed Journal: Eur J Anaesthesiol ISSN: 0265-0215 Impact factor: 4.330