Potential problems with selective pulses in NMR imaging systems.

P M Joseph, L Axel, M O'Donnell.   

Abstract

Most nuclear magnetic resonance (NMR) imaging systems require pulses whose frequency spectrum is shaped so as to selectively excite a given plane in the presence of a magnetic field gradient. We demonstrate by both computer simulation and experiment that linear Fourier transform theory is not a reliable guide to the uniformity of flip angle in the slice. We show by simulation that the nonuniformity can have serious implications for the measurement of relaxation time T1 if selective 180 degrees pulses are used; the exact results depend also on the details of data analysis and criteria for adjusting the rephasing gradients. We describe an experiment and a phantom in which the axial nonuniformity can be demonstrated on clinical NMR imaging machines.

Mesh:

Year:  1984        PMID: 6513885     DOI: 10.1118/1.595579

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

1.  PROPELLER EPI: an MRI technique suitable for diffusion tensor imaging at high field strength with reduced geometric distortions.

Authors:  Fu-Nien Wang; Teng-Yi Huang; Fa-Hsuan Lin; Tzu-Chao Chuang; Nan-Kuei Chen; Hsiao-Wen Chung; Cheng-Yu Chen; Kenneth K Kwong
Journal:  Magn Reson Med       Date:  2005-11       Impact factor: 4.668

2.  MR ventriculocisternography by using 3D balanced steady-state free precession imaging: technical note.

Authors:  Ming-Long Wu; Cheng-Wen Ko; Ting-Yi Chen; Ming-Ting Wu; Hsiao-Wen Chung; Teng-Yi Huang; Yi-Ru Lin
Journal:  AJNR Am J Neuroradiol       Date:  2005-05       Impact factor: 3.825

3.  Minimum acquisition methods for simultaneously imaging T(1), T(2), and proton density with B(1) correction and no spin-echoes.

Authors:  Guan Wang; AbdEl-Monem M El-Sharkawy; Paul A Bottomley
Journal:  J Magn Reson       Date:  2014-03-01       Impact factor: 2.229

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.