| Literature DB >> 649767 |
.
Abstract
The performance of autoclaves in 27 laboratories, operated in accordance with the normal routine of local practice, has been monitored using thermometric equipment. Sterilising performance was unsatisfactory on 10 of 62 occasions, and cooling was inadequate on 52 of 60 occasions.Mesh:
Year: 1978 PMID: 649767 PMCID: PMC1145294 DOI: 10.1136/jcp.31.5.418