Autoclaves and their dangers and safety in laboratories.

E H Gillespie, S A Gibbons.   

Abstract

Using a laboratory downward displacement vertical autoclave with the help of thermocouples recorded on a 12 point multichannel strip recorder, the risk of failing to sterilize laboratory discard buckets has been demonstrated. The use of proper temperature and time controls can prevent this risk. A load in a bucket with perforated sides is more easily sterilized than in a solid bucket. Wire baskets, where appropriate, facilitate the sterilizing practice. The addition of water to a bucket does not reduce the time of heating up. It is desirable that sealed bottles of media should not be sterilized in simple downward displacement autoclaves, but if used, strict monitoring of temperatures and times is essential both in the heating up stage and especially in the cooling stage. The temperatures in bottles are slow to rise and very slow to fall. Bottles at high temperature 80-105 degrees C. or over have a high internal pressure which can allow the bottles to explode when subjected to thermal shock if removed too early. It is suggested that all laboratory autoclaves should have a load temperature simulator or similar device to control the temperature of the load during the cycle automatically. For the sterilization of fluid media, it is suggested that, in addition to a simulator there should be accelerated cooling to reduce damage to the media and, what is more important, to rapidly bring down the temperature and thus the internal pressure in the bottles to a safe level. The opening of the sterilizer door or lid should be automatically controlled by the load temperature simulator.

Entities:  

Mesh:

Substances:

Year:  1975        PMID: 1059711      PMCID: PMC2130361          DOI: 10.1017/s0022172400024517

Source DB:  PubMed          Journal:  J Hyg (Lond)        ISSN: 0022-1724


  5 in total

1.  Influence of packaging and processing conditions on the decontamination of laboratory biomedical waste by steam sterilization.

Authors:  G Ozanne; R Huot; C Montpetit
Journal:  Appl Environ Microbiol       Date:  1993-12       Impact factor: 4.792

2.  Sterilisation in the laboratory autoclave using direct air displacement by steam.

Authors:  P H Everall; C A Morris; R Yarnell
Journal:  J Clin Pathol       Date:  1978-02       Impact factor: 3.411

3.  Autoclaving practice in microbiology laboratories: report of a survey. The Public Health Laboratory Service Subcommittee on laboratory autoclaves.

Authors: 
Journal:  J Clin Pathol       Date:  1978-05       Impact factor: 3.411

4.  Isolation of Shigella sonnei by fluid media.

Authors:  T H Price
Journal:  J Hyg (Lond)       Date:  1976-12

5.  Decontamination of laboratory microbiological waste by steam sterilization.

Authors:  W A Rutala; M M Stiegel; F A Sarubbi
Journal:  Appl Environ Microbiol       Date:  1982-06       Impact factor: 4.792

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.