D L McDaniel, G Cohen, L K Wagner, L H Robinson.
Abstract
The relative dose efficiencies (RDE) of various antiscatter grids and air gaps were determined for conditions simulating those found in pediatric radiography, using phantoms representing a newborn child, a 5-yr-old and a 10-yr-old child. Our data indicate than an air gap is best for the newborn, due to the low levels of scatter. The 8:1 fiber grid or 15.2-cm air gap without a grid can improve dose efficiency (DE) for the 5-yr-old child by 20%-25% relative to the 3.3-cm air gap and no-grid technique, while for the 10-yr-old child, DE can be improved by 40% with an 8:1 fiber grid.Entities: