| Literature DB >> 6478694 |
R R Jacobs, F Schlaepfer, R Mathys, A Nachemson, S M Perren.
Abstract
Existing internal fixation systems for the injured or deformed spine present problems with overdistraction and control of the contoured rod necessary for transverse forces. A locking hook spinal rod avoids these problems by using a locking cover to secure the lamina in the hook and meshing radial grooves to lock the contoured rod to both the upper and lower hooks in 6 degrees intervals of rotation. The 7-mm stainless-steel rod is 50% stronger than the 1/4-in Harrington rod and also avoids the weakening effect of the notches. Cadaver spine testing gives nearly a threefold increase in failure strength (125 +/- 17 Nm versus 44.1 +/- 2.1).Entities:
Mesh:
Year: 1984 PMID: 6478694