Literature DB >> 645714

Rapid miniaturized chromatographic quality control procedures for iodinated radiopharmaceuticals.

A M Zimmer, D G Pavel.   

Abstract

A rapid, miniaturized chromatographic system was developed to determine the labeling efficiency of iodinated radio-pharmaceuticals including I-131 orthoiodohippurate, I-131 rose bengal, I-125 RISA, and I-131 19-iodocholesterol. The system consists of three specific chromatography strips (Whatman 31ET paper chromatography, silica gel and silicic acid instant thin-layer chromatography) and three solvents (chloroform:glacial acetic acid in a 1.00:0.05 v/v mixture, absolute methanol, and 0.9% sodium chloride). The strips are color-coded, miniaturized, and appropriately labeled. The chromatographic quality control procedures are easy to use, rapid, and can be incorporated easily in a routine quality control program.

Entities:  

Mesh:

Substances:

Year:  1978        PMID: 645714

Source DB:  PubMed          Journal:  Am J Hosp Pharm        ISSN: 0002-9289


  1 in total

1.  Rapid miniaturized chromatography for Tc-99m IDA agents: comparison with gel chromatography.

Authors:  A M Zimmer; W Majewski; S M Spies
Journal:  Eur J Nucl Med       Date:  1982