Calibration of some Wright peak flow meters.

J Fisher, A Shaw.   

Abstract

The performance of Wright peak flow meters is described using an absolute and precise calibration method. This showed errors as great as 25% with the peak flow meter and 40% with the mini peak flow meter. At large flow rates errors were less. The expected variation between individual peak flow meters was about 12%. The expected variation between individual mini peak flow meters was between 16% and 22%, depending on flow rate.

Mesh:

Year:  1980        PMID: 6445741     DOI: 10.1093/bja/52.4.461

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166