| Literature DB >> 6437002 |
D Davis, H W Bonekat, D Andrews, J W Shigeoka.
Abstract
In view of recent reports of contamination of the fibreoptic bronchoscope by tuberculous and non-tuberculous mycobacteria, we evaluated the disinfecting properties of aqueous glutaraldehyde (Cidex) and an iodophor (Prepodyne) against Mycobacterium tuberculosis and M gordonae. We found that a 15 minute disinfection procedure with either agent, coupled with initial vigorous mechanical cleaning of the bronchoscope and its accessories, is a quick and reliable method for preventing the contamination of the bronchoscope with mycobacteria.Entities:
Mesh:
Substances:
Year: 1984 PMID: 6437002 PMCID: PMC459919 DOI: 10.1136/thx.39.10.785