Detection of arrhythmia: limited usefulness of patient activated recording devices.

L E Thomas, L M Shapiro, E J Perrins, K M Fox.   

Abstract

The findings of 24 hour ambulatory electrocardiography and monitoring with the Cardiomemo, a device recording 32 seconds of electrocardiogram during symptoms, were compared in 20 patients with symptoms suggestive of arrhythmia. Ambulatory electrocardiography showed arrhythmia in seven patients, extrasystoles in six, and normal findings in seven. Nine patients failed to transmit any Cardiomemo recordings, and the Cardiomemo failed to show ventricular and supraventricular tachycardia. It did not show any appreciable arrhythmia in the seven patients with normal 24 hour electrocardiograms. The Cardiomemo does not offer any important advantages over ambulatory electrocardiography, and its relative cheapness is outweighed by the limited number of patients who can use the device in one year. It can, however, reassure anxious patients of the absence of arrhythmia during symptoms.

Entities:  

Mesh:

Year:  1984        PMID: 6435794      PMCID: PMC1443234          DOI: 10.1136/bmj.289.6452.1106

Source DB:  PubMed          Journal:  Br Med J (Clin Res Ed)        ISSN: 0267-0623


  3 in total

1.  Evaluation of outpatient arrhythmias utilizing transtelephonic monitoring.

Authors:  P Judson; D R Holmes; W P Baker
Journal:  Am Heart J       Date:  1979-06       Impact factor: 4.749

2.  Current status of ambulatory electrocardiography.

Authors:  R A Winkle
Journal:  Am Heart J       Date:  1981-10       Impact factor: 4.749

3.  Arrhythmia surveillance by transtelephonic monitoring: comparison with Holter monitoring in symptomatic ambulatory patients.

Authors:  R S Grodman; R J Capone; A S Most
Journal:  Am Heart J       Date:  1979-10       Impact factor: 4.749

  3 in total
  2 in total

1.  Transtelephonic electrocardiographic monitoring for detection and treatment of cardiac arrhythmia.

Authors:  A F Safe; R T Maxwell
Journal:  Postgrad Med J       Date:  1990-02       Impact factor: 2.401

2.  Detection of arrhythmias: use of a patient-activated ambulatory electrocardiogram device with a solid-state memory loop.

Authors:  A P Brown; K D Dawkins; J G Davies
Journal:  Br Heart J       Date:  1987-09
  2 in total