I Hemmilä, S Dakubu, V M Mukkala, H Siitari, T Lövgren.
Abstract
A nonisotopic immunoassay has been developed based on a sensitive detection of europium (III) in water solution using time-resolved fluorometry. The europium label is bound to the antibody with EDTA derivatives, either diazophenyl-EDTA-Eu or isothiocyanatophenyl-EDTA-Eu. After the immunometric assay has been completed the europium is preferably dissociated from the antibody at low pH and measured by time-resolved fluorescence in a micellar solution containing Triton X-100, beta-diketone, and a Lewis base. The detergent solubilizes the chelating compounds in the solution and excludes water from the fluorescent ligand-europium complex. Europium concentrations as low as 5 X 10(-14)M were measured using a 1-s counting time. The sensitivity of the immunoassay of rabbit IgG used as a model system was 25 pg/ml (6 pg/assay).Entities:
Mesh:
Substances:
Year: 1984 PMID: 6375455 DOI: 10.1016/0003-2697(84)90095-2
Source DB: PubMed Journal: Anal Biochem ISSN: 0003-2697 Impact factor: 3.365