Physical and clinical evaluation of new high-strip-density radiographic grids.

K Doi, P H Frank, H P Chan, C J Vyborny, S Makino, N Iida, M Carlin.   

Abstract

The imaging performance of new high-strip-density (HSD) grids having 57 lines/cm was compared with that of conventional low-strip-density (LSD) grids having 33 or 40 lines/cm. The unique advantage of HSD grids is that, under most standard radiographic conditions, the grid lines are not noticeable on the final image, even if the grid is stationary. This is due to the combined effect of the high fundamental spatial frequency of HSD grids, the modulation transfer function of screen-film systems and of the human visual system, and scattered radiation. Monte Carlo simulation studies, phantom images, and clinical evaluation indicate that HSD grids can provide contrast improvement factors and Bucky factors that are comparable to or slightly better than those obtained with LSD grids. Therefore, it may now be possible to eliminate moving Bucky trays from radiographic tables and fluoroscopic devices.

Entities:  

Mesh:

Year:  1983        PMID: 6340160     DOI: 10.1148/radiology.147.2.6340160

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  1 in total

1.  Investigation of optimum anti-scatter grid selection for digital radiography: physical imaging properties and detectability of low-contrast signals.

Authors:  Nobukazu Tanaka; Kentaro Naka; Aya Saito; Junji Morishita; Fukai Toyofuku; Masafumi Ohki; Yoshiharu Higashida
Journal:  Radiol Phys Technol       Date:  2012-08-08