Ten-year pathological study of Beall mitral valve prostheses.

K S Chang, J Fernandez, S D Cha, S S Yang.   

Abstract

Detailed pathological studies are reported on 20 cloth-covered Beall mitral valve prostheses, which had to be removed for valve failure nine months to 10 years after implantation. Results in all patients except one indicated that there was a progressive and persistent granulomatous reaction that continued up to at least 10 years. Although the rate of tissue ingrowth into the valve decreased with time, the total volume of tissue formation increased. Cloth wear and disruption and deformity of the teflon discs was frequently seen. In these 20 valves the desired result--namely, the complete and smooth endothelialisation of the prosthesis--was certainly not achieved.

Entities:  

Mesh:

Substances:

Year:  1982        PMID: 6210973      PMCID: PMC459265          DOI: 10.1136/thx.37.2.137

Source DB:  PubMed          Journal:  Thorax        ISSN: 0040-6376            Impact factor:   9.139


  11 in total

1.  The pathology of wear in the Beall model 104 heart valve prosthesis.

Authors:  M D Silver; G J Wilson
Journal:  Circulation       Date:  1977-10       Impact factor: 29.690

2.  Results of mitral valve replacement with the Beall prosthesis in 209 patients.

Authors:  J Fermandez; D Morse; P Spagna; G Lemole; A Gooch; S S Yang; V Maranhao
Journal:  J Thorac Cardiovasc Surg       Date:  1976-02       Impact factor: 5.209

3.  The granulomatous inflammatory response. A review.

Authors:  D O Adams
Journal:  Am J Pathol       Date:  1976-07       Impact factor: 4.307

Review 4.  Participation of mononuclear phagocytes in chronic inflammatory diseases.

Authors:  R C Page; P Davies; A C Allison
Journal:  J Reticuloendothel Soc       Date:  1974-05

Review 5.  The granulomatous inflammatory exudate.

Authors:  W G Spector
Journal:  Int Rev Exp Pathol       Date:  1969

6.  Improved results in 336 patients with the isolated mitral Beall valve replacement.

Authors:  H T Nichols; J Fernandez; D Morse; A S Gooch
Journal:  Chest       Date:  1972-09       Impact factor: 9.410

7.  Tissue ingrowth and the rigid heart valve. Review of clinical and experimental experience during the past year.

Authors:  N S Braunwald; A G Morrow
Journal:  J Thorac Cardiovasc Surg       Date:  1968-09       Impact factor: 5.209

8.  Long-term behavior of a dacron arterial substitute: clinical, roentgenologic and histologic correlations.

Authors:  D E Szilagyi; R F Smith; J P Elliott; H M Allen
Journal:  Ann Surg       Date:  1965-09       Impact factor: 12.969

9.  The prevention of thrombosis in artificial cardiac valves.

Authors:  J C Davila; F Amongero; R S Sethi; N L Rincon; T E Palmer; E V Lautsch
Journal:  Ann Thorac Surg       Date:  1966-09       Impact factor: 4.330

10.  Sarcoma associated with dacron prosthetic material: case report and review of the literature.

Authors:  T X O'Connell; H J Fee; A Golding
Journal:  J Thorac Cardiovasc Surg       Date:  1976-07       Impact factor: 5.209