Literature DB >> 6181468

Myopotential inhibition of unipolar AV sequential (DVI) pacemaker.

H J Echeverria, R M Luceri, R J Thurer, A Castellanos.   

Abstract

Whereas myopotential inhibition of QRS-inhibited (VVI) pacemakers is well known, its occurrence in patients with AV sequential (DVI) pacemakers has not been reported. The present communication deals with spontaneous and induced myopotential inhibition of a multiprogrammable Intermedics unipolar AV sequential (DVI) pacemaker. The bedside maneuvers that were performed in the patient exposed the problem, therefore serving to establish the diagnosis. Although external adjustment of the sensitivity was the simple, non-invasive solution in this case, more studies are required to determine the success rate of this approach as well as the incidence and clinical significance of myopotential inhibition of unipolar DVI pacemakers.

Entities:  

Mesh:

Year:  1982        PMID: 6181468     DOI: 10.1111/j.1540-8159.1982.tb02186.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  1 in total

1.  Prevention of myopotential inhibition of unipolar QRS-inhibited demand pacemakers.

Authors:  H Aida
Journal:  Jpn J Surg       Date:  1983-11