Myopotential inhibition of unipolar QRS-inhibited (VVI) pacemakers, assessed by ambulatory Holter monitoring of the electrocardiogram.

K Breivik, O J Ohm.   

Abstract

Seventy-four patients with unipolar QRS-inhibited pacemakers (VVI) were Holter monitored to assess the occurrence of pacemaker inhibition caused by skeletal muscle potentials during daily activities. Fifty patients had high-grade atrioventricular block and 24 had sinoatrial disease. Chest wall stimulation prior to monitoring revealed asystole of > 4 seconds duration in 22 patients, and ventricular rates between 25 and 56 beats per minute in 52 patients. Fifty-one patients (69%) had one or more episodes of pacemaker inhibition from myopotentials. Inhibition occurred in all types of pacemakers studies, but was most frequent and of longest duration in patients with Siemens-Elema 207/70 (13/14 patients), Cordis Omni-Stanicor (6/7 patients), CPI Microlith (5/6 patients), and Medtronic 5945 (8/10 patients). This was not unexpected considering the filter characteristics of the pacemakers. Nine patients (12%) presented symptoms which might be ascribed to pacemaker inhibition. The longest asystole observed was 3.2 s. Seven patients had pacemakers spikes falling on or near T-waves of spontaneous heart beats because their pacemakers had been rendered refractory by myopotentials. No serious arrhythmias were seen during episodes of pacemaker inhibition or interference. More emphasis should be put on the improvement of filter characteristics of unipolar VVI-pacemakers. Pacemaker patients with symptoms of myopotential inhibition should be equipped with either a bipolar or ventricular triggered (VVT) pacemaker or with a sensitivity and/or pacing mode programmable pacemaker.

Entities:  

Mesh:

Year:  1980        PMID: 6160541     DOI: 10.1111/j.1540-8159.1980.tb05257.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  5 in total

1.  Myopotential inhibition of demand pacemakers.

Authors:  A V Lazar; M J Mihalick; S A De Young; R J Hall
Journal:  Tex Heart Inst J       Date:  1982-06

2.  Effects of muscle potential depression and muscle stimulation caused by different insulation coating configurations on cardiac pacemakers.

Authors:  Toshimi Yajima; Kenichi Yamada; Naoko Okubo; Takashi Nitta; Masami Ochi; Kazuo Shimizu
Journal:  J Artif Organs       Date:  2005       Impact factor: 1.731

3.  Prevention of myopotential inhibition of unipolar QRS-inhibited demand pacemakers.

Authors:  H Aida
Journal:  Jpn J Surg       Date:  1983-11

4.  Myopotential inhibition of unipolar pacemakers.

Authors:  D R Ramsdale; R G Charles
Journal:  Br Med J (Clin Res Ed)       Date:  1983-11-12

5.  Environmental hazards for pacemaker patients.

Authors:  E Sowton
Journal:  J R Coll Physicians Lond       Date:  1982-07
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.