Elimination of toxicity from polyvinyl trays after sterilization with ethylene oxide.

A Rueter, J B Schleicher.   

Abstract

Ethylene oxide (ETO) sterilization of polyvinyl trays used in an antiviral screening program was initiated to overcome seasonal outbreaks of bacterial and mycotic contamination of tissue culture cells. Trays sterilized by 100% ETO for 5 hr after 48 hr of prehumidification, or by 12% ETO plus 88% Freon 12 for 1, 3, 5 or 18 hr, followed by various methods of aeration, were seeded with several types of tissue culture cells and examined for contamination, toxicity, and monolayer quality. A 1-hr exposure in 12% ETO plus 88% Freon 12 was adequate for sterilization, although residual toxicity for tissue cultures remained. A 7-day aeration period at 37 C was sufficient to eliminate toxicity and allow the growth of good monolayers of WI-38, HEp-2 and primary bovine kidney cells. Sterilization with 100% ETO required 14 days of aeration at 37 C to eliminate cytotoxicity. Increased residual toxicity resulting from longer ETO sterilization periods required longer aeration times at 37 C or higher aeration temperatures for detoxification.

Entities:  

Mesh:

Substances:

Year:  1969        PMID: 5392607      PMCID: PMC378192          DOI: 10.1128/am.18.6.1057-1059.1969

Source DB:  PubMed          Journal:  Appl Microbiol        ISSN: 0003-6919


  2 in total

1.  Toxicological studies on certain medical grade plastics sterilized by ethylene oxide.

Authors:  R K O'Leary; W L Guess
Journal:  J Pharm Sci       Date:  1968-01       Impact factor: 3.534

2.  [Experimental studies of a new sterilization method using ethylene oxide under partial vacuum].

Authors:  A L Brunel; F Archinet; R Vannier
Journal:  Rev Stomatol Chir Maxillofac       Date:  1967 Apr-Mar
  2 in total
  3 in total

1.  Biologic hernia implants in experimental intraperitoneal onlay mesh plasty repair: the impact of proprietary collagen processing methods and fibrin sealant application on tissue integration.

Authors:  A H Petter-Puchner; R H Fortelny; K Silic; J Brand; S Gruber-Blum; H Redl
Journal:  Surg Endosc       Date:  2011-04-30       Impact factor: 4.584

2.  Two cross-linked porcine dermal implants in a single patient undergoing hernia repair.

Authors:  Luke A Linz; Leandra H Burke; Lisa A Miller
Journal:  BMJ Case Rep       Date:  2013-01-22

3.  Disinfecting endoscopes: how not to transmit Mycobacterium tuberculosis by bronchoscopy.

Authors:  W D Leers
Journal:  Can Med Assoc J       Date:  1980-08-23       Impact factor: 8.262

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.