Literature DB >> 534707

Failure of conventional monitoring to detect apnea resulting in hypoxemia.

J L Peabody, G A Gregory, M M Willis, A G Philip, J F Lucey.   

Abstract

In summary, our findings support and confirm the concerns of many investigators that present methods of cardiorespiratory monitoring are inadequate for the detection of many forms of apnea. Nurses underrecord both the frequency and duration of apneic episodes. Bradycardia is an unreliable index of hypoxemia. Thoracic impedance monitors are unreliable because they detect only a fixed duration of respiratory pause and are sensitive to many artifacts unavoidable in a clinical setting. Finally, ineffective breathing patterns such as disorganized breathing, obstructive apnea, and paradoxical breathing are undetectable by thoracic impedance monitoring. We warn against the reliance on heart rate and thoracic impedance monitoring alone for infants with recurrent apnea.

Entities:  

Mesh:

Substances:

Year:  1979        PMID: 534707

Source DB:  PubMed          Journal:  Birth Defects Orig Artic Ser        ISSN: 0547-6844


  6 in total

1.  Does caffeine prevent hypoxaemic episodes in premature infants? A randomized controlled trial.

Authors:  H U Bucher; G Duc
Journal:  Eur J Pediatr       Date:  1988-04       Impact factor: 3.183

2.  Effects of routine care procedures on transcutaneous oxygen in neonates: a quantitative approach.

Authors:  D A Danford; S Miske; J Headley; R M Nelson
Journal:  Arch Dis Child       Date:  1983-01       Impact factor: 3.791

3.  Very long apnea events in preterm infants.

Authors:  Mary A Mohr; Brooke D Vergales; Hoshik Lee; Matthew T Clark; Douglas E Lake; Anne C Mennen; John Kattwinkel; Robert A Sinkin; J Randall Moorman; Karen D Fairchild; John B Delos
Journal:  J Appl Physiol (1985)       Date:  2014-12-30

4.  Anemia, apnea of prematurity, and blood transfusions.

Authors:  Kelley Zagol; Douglas E Lake; Brooke Vergales; Marion E Moorman; Alix Paget-Brown; Hoshik Lee; Craig G Rusin; John B Delos; Matthew T Clark; J Randall Moorman; John Kattwinkel
Journal:  J Pediatr       Date:  2012-04-10       Impact factor: 4.406

Review 5.  Transcutaneous PO2 measurement.

Authors:  K K Tremper
Journal:  Can Anaesth Soc J       Date:  1984-11

6.  Accurate automated apnea analysis in preterm infants.

Authors:  Brooke D Vergales; Alix O Paget-Brown; Hoshik Lee; Lauren E Guin; Terri J Smoot; Craig G Rusin; Matthew T Clark; John B Delos; Karen D Fairchild; Douglas E Lake; Randall Moorman; John Kattwinkel
Journal:  Am J Perinatol       Date:  2013-04-16       Impact factor: 1.862