Literature DB >> 519961

Impairment of flow in routine gravity-fed intravenous infusions to surgical patients.

E W Clarke, J P Jamison, J B Quartey-Papafio.   

Abstract

1. Gravity-fed intravenous infusions were studied in patients by using continuous flow recording and a new resistance-measuring method. 2. A common, though preventable, cause of reduced flow was the gradual deformation of the plastic tubing under the regulator clamp. 3. In many cases there was some constriction of the cannulated vein and in a few this constriction was severe enough to seriously impair the infusion. The severity of the constriction varied periodically. 4. There was no evidence of troble due to a generally raised pressure in the arm veins nor was clotting in the cannula a serious problem. 5. A simple test is described which may be found useful clinically for assessing the difficult infusion.

Entities:  

Mesh:

Year:  1979        PMID: 519961     DOI: 10.1042/cs0570515

Source DB:  PubMed          Journal:  Clin Sci (Lond)        ISSN: 0143-5221            Impact factor:   6.124


  1 in total

1.  An analysis of the effect of venous resistance on the performance of gravity-fed intravenous infusion systems.

Authors:  D B Goodie; J H Philip
Journal:  J Clin Monit       Date:  1994-07
  1 in total