Contact pressure dependence of wear rates of ultra high molecular weight polyethylene.

W Rostoker, J O Galante.   

Abstract

Wear rates of orthopaedic polyethylene in 37 degrees C water are not very contact stress dependent below 1000 psi (6.9 MPa) but above that level they accelerate substantially. The pressure dependence overall follows an exponential function. Creep in the contact pressure range of 3--17 MPa and above is a much larger factor than wear in indentation effects. For accurate measurement of wear rates by depth measurements it is necessary to permit creep to proceed under static load until it ceases before beginning the wear test.

Entities:  

Mesh:

Substances:

Year:  1979        PMID: 511863     DOI: 10.1002/jbm.820130612

Source DB:  PubMed          Journal:  J Biomed Mater Res        ISSN: 0021-9304


  6 in total

1.  The influence of stress conditions on the wear of UHMWPE for total joint replacements.

Authors:  P S Barbour; D C Barton; J Fisher
Journal:  J Mater Sci Mater Med       Date:  1997-10       Impact factor: 3.896

2.  [AMC unicondylar prosthesis].

Authors:  G Saxler; D Temmen; G Bontemps
Journal:  Orthopade       Date:  2004-11       Impact factor: 1.087

Review 3.  Advances in tribological testing of artificial joint biomaterials using multidirectional pin-on-disk testers.

Authors:  D Baykal; R S Siskey; H Haider; V Saikko; T Ahlroos; S M Kurtz
Journal:  J Mech Behav Biomed Mater       Date:  2013-05-29

4.  Wear performance of inverted non-conforming bearings in anatomic total shoulder arthroplasty.

Authors:  Lorenzo Banci; Alessio Meoli; Martin Hintner; Hans Rudolf Bloch
Journal:  Shoulder Elbow       Date:  2019-02-07

5.  The effect of geometry and abduction angle on the stresses in cemented UHMWPE acetabular cups--finite element simulations and experimental tests.

Authors:  Rami K Korhonen; Arto Koistinen; Yrjö T Konttinen; Seppo S Santavirta; Reijo Lappalainen
Journal:  Biomed Eng Online       Date:  2005-05-17       Impact factor: 2.819

6.  Patient and surgery related factors associated with fatigue type polyethylene wear on 49 PCA and DURACON retrievals at autopsy and revision.

Authors:  Markus Rohrbach; Martin Lüem; Peter E Ochsner
Journal:  J Orthop Surg Res       Date:  2008-02-22       Impact factor: 2.359

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.