Improved accuracy of continuous measurement of arterial oxygen tension in sick newborn infants. Criteria for reading and recalibrating the electrode.

A R Wilkinson, R H Phibbs, G A Gregory.   

Abstract

Improved accuracy of continuous arterial oxygen tension (PaO2) measurement is obtained if allowance is made for the response time of the electrode. Recalibration is necessary only when PaO2 measured in a blood sample differs from the electrode value by more than +/- 11.5 mmHg (1.5 kPa).

Entities:  

Mesh:

Substances:

Year:  1979        PMID: 453914      PMCID: PMC1545289          DOI: 10.1136/adc.54.4.307

Source DB:  PubMed          Journal:  Arch Dis Child        ISSN: 0003-9888            Impact factor:   3.791


  5 in total

1.  Use of a continuously recording intravascular oxygen electrode in the newborn.

Authors:  P Goddard; I Keith; H Marcovitch; N R Roberton; P Rolfe; J W Scopes
Journal:  Arch Dis Child       Date:  1974-11       Impact factor: 3.791

2.  Catheter-tip transducer for continuous in-vivo measurement of oxygen tension.

Authors:  D Parker; A Key; R S Davies
Journal:  Lancet       Date:  1971-05-08       Impact factor: 79.321

3.  Complications of umbilical oxygen electrodes.

Authors:  K R Kollmeyer; R C Tsang
Journal:  J Pediatr       Date:  1974-06       Impact factor: 4.406

Review 4.  Catheterization of umbilical vessels in newborn infants.

Authors:  J A Kitterman; R H Phibbs; W H Tooley
Journal:  Pediatr Clin North Am       Date:  1970-11       Impact factor: 3.278

5.  Continuous arterial oxygen tension monitoring in the newborn infant.

Authors:  T R Harris; M Nugent
Journal:  J Pediatr       Date:  1973-06       Impact factor: 4.406

  5 in total
  2 in total

1.  Pulse oximetry in preterm infants.

Authors:  A Wasunna; A G Whitelaw
Journal:  Arch Dis Child       Date:  1987-09       Impact factor: 3.791

2.  Continuous neonatal blood gas monitoring using a multiparameter intra-arterial sensor.

Authors:  C Morgan; S J Newell; D A Ducker; J Hodgkinson; D K White; C J Morley; J M Church
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  1999-03       Impact factor: 5.747

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.