D Pitkin, P Actor, W Holl, A Post, J A Weisbach.
Abstract
The Autoturb System, a semiautomated system for photometric bioassay, was used to determine cefazolin content. Suitable conditions for the assay using Streptococcus faecalis ATCC 10541 as the indicator organism included a medium pH of 6.0 to 7.0 and an incubation time of 3 to 3.5 h at 36 C. Multiple independent assays of samples from a common batch showed the test to be highly reproducible. Accuracy of the turbidimetric assay was evaluated by comparing data obtained from a chemical (hydroxylamine) and a biological (disk diffusion) assay. The available data show the turbidimetric assay to be a rapid, accurate, and reproducible method for determining the biological activity of cefazolin samples.Entities:
Mesh:
Substances:
Year: 1974 PMID: 4209068 PMCID: PMC428952 DOI: 10.1128/AAC.5.3.223
Source DB: PubMed Journal: Antimicrob Agents Chemother ISSN: 0066-4804 Impact factor: 5.191