| Literature DB >> 407434 |
C L Wingate, J O Archambeau, A M Koehler, G W Bennett.
Abstract
Accurate bolus is needed for extension of Bragg-peak therapy. Proton beam-stopping profiles in a lucite-styrofoam-tissue phantom and in a Rando phantom were recorded photographically. Air volumes caused the largest distortions. Lucite bolus was cut to achieve desired beam-stopping profiles. Verification of bolus effect in situ will be important to control beam penetration within 5 mm.Mesh:
Substances:
Year: 1977 PMID: 407434 DOI: 10.1118/1.594365