Accuracy of two ear oximeters at rest and during exercise in pulmonary patients.

A L Ries, J T Farrow, J L Clausen.   

Abstract

The accuracy of 2 ear oximeters (Hewlett-Packard 47201A and Biox IIA) was evaluated at rest and during exercise in 116 pulmonary patients. Ear oximetry measurements were compared with arterial oxygen saturation (SaO2) measured from simultaneous arterial blood samples. When used according to the manufacturer's instructions for stabilization during exercise testing, both oximeters demonstrated comparable accuracy in estimating arterial blood SaO2 both at rest and during exercise in these patients (95% confidence limits, 2 SEE = +/- 4 to 5%). The oximeters were more accurate in measuring a change in SaO2 from a previous reading (95% confidence limits, 2 SEE = +/- 2.5 to 3.5%). COHgb levels greater than 4% and dark skin pigmentation appeared to decrease the accuracy of ear oximetry measurements. Within the limitations of accuracy, both ear oximeters may be useful for clinical exercise testing in pulmonary patients.

Entities:  

Mesh:

Substances:

Year:  1985        PMID: 4037542     DOI: 10.1164/arrd.1985.132.3.685

Source DB:  PubMed          Journal:  Am Rev Respir Dis        ISSN: 0003-0805


  12 in total

Review 1.  Pulse oximetry and capnography in intensive and transitional care units.

Authors:  F Bongard; D Sue
Journal:  West J Med       Date:  1992-01

Review 2.  Exercise testing in the assessment of pulmonary disease.

Authors:  S P Blackie; R L Pardy
Journal:  Clin Rev Allergy       Date:  1990 Summer-Fall

Review 3.  Pulse oximetry versus arterial blood gas specimens in long-term oxygen therapy.

Authors:  D J Pierson
Journal:  Lung       Date:  1990       Impact factor: 2.584

4.  Hypoxaemia during chest physiotherapy in patients with cystic fibrosis.

Authors:  T McDonnell; W T McNicholas; M X FitzGerald
Journal:  Ir J Med Sci       Date:  1986-10       Impact factor: 1.568

5.  Welcome the new era of oximetry.

Authors:  R J Fallat
Journal:  Intensive Care Med       Date:  1988       Impact factor: 17.440

6.  Incidence of exercise induced hypoxemia in elite endurance athletes at sea level.

Authors:  S K Powers; S Dodd; J Lawler; G Landry; M Kirtley; T McKnight; S Grinton
Journal:  Eur J Appl Physiol Occup Physiol       Date:  1988

7.  Continuous dual oximetry in surgical critical care. Indications and limitations.

Authors:  F S Bongard; T A Leighton
Journal:  Ann Surg       Date:  1992-07       Impact factor: 12.969

8.  Influence of Sex, BMI, and Skin Color on the Accuracy of Non-Invasive Cuffless Photoplethysmography-Based Blood Pressure Measurements.

Authors:  Dean Nachman; Arik Eisenkraft; Nir Goldstein; Arik Ben-Ishay; Meir Fons; Roei Merin; Yftach Gepner
Journal:  Front Physiol       Date:  2022-06-29       Impact factor: 4.755

9.  The accuracy of the Biox 3700 pulse oximeter in patients receiving vasoactive therapy.

Authors:  J Ibáñez; J Velasco; J M Raurich
Journal:  Intensive Care Med       Date:  1991       Impact factor: 17.440

10.  Re-examination of the incidence of exercise-induced hypoxaemia in highly trained subjects.

Authors:  D D Brown; R G Knowlton; P B Sanjabi; B T Szurgot
Journal:  Br J Sports Med       Date:  1993-09       Impact factor: 13.800

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.