Literature DB >> 4037476

Effects of rapid infusion with high pressure and large-bore i.v. tubing on red blood cell lysis and warming.

J R Mateer, B W Perry, B M Thompson, J F Tucker, C Aprahamian.   

Abstract

A prototype large-bore intravenous tubing was developed and tested. Mean flow rates for blood (Hct 45%) and tap water were determined for several catheters at 600 mm Hg, 300 mm Hg, and gravity flow and were statistically analyzed by calculating the 95% confidence intervals. The degree of hemolysis during high pressure and flow was determined by measuring the plasma free hemoglobin using the spectrophotometric method. To determine if cold banked blood can be adequately warmed at high flow rates, thermocouples were used to measure the blood temperature before and after rapid infusion through a blood warmer. Results included maximum flow rates of 1,764 mL/min for tap water, and 1,714 mL/min for blood (Hct 45%) at 600 mm Hg through the large-bore tubing and an 8.5-F catheter. Flow rates for other pressure and catheter combinations were tabulated. The plasma-free hemoglobin increased slightly compared to controls with high pressure (less than or equal to 600 mm Hg) and flow rates. The increase correlated with less than 1% red blood cell lysis in all trials. When 13 C blood was infused through a warmer, blood temperature increased to 25.3 C at the maximum flow rate of 732 mL/min. Slightly higher heat gain resulted with slower infusion rates. We conclude that the prototype large-bore tubing and up to 600 mm Hg pressure provide rapid flow rates without significant hemolysis. Blood warming may be inadequate at higher flow rates.

Entities:  

Mesh:

Year:  1985        PMID: 4037476     DOI: 10.1016/s0196-0644(85)80238-9

Source DB:  PubMed          Journal:  Ann Emerg Med        ISSN: 0196-0644            Impact factor:   5.721


  4 in total

1.  Magnetron-based inline microwave fluid warmer.

Authors:  G L Gibby; S Lampotang; N Gravenstein
Journal:  J Clin Monit       Date:  1996-05

2.  In-line pressure within a HOTLINE® Fluid Warmer, under various flow conditions.

Authors:  Midoriko Higashi; Ken Yamaura; Yukie Matsubara; Takuya Fukudome; Sumio Hoka
Journal:  J Clin Monit Comput       Date:  2014-08-03       Impact factor: 2.502

3.  Pressure Infusion Cuff and Blood Warmer during Massive Transfusion: An Experimental Study About Hemolysis and Hypothermia.

Authors:  Thomas G Poder; Denise Pruneau; Josée Dorval; Louis Thibault; Jean-François Fisette; Suzanne K Bédard; Annie Jacques; Patrice Beauregard
Journal:  PLoS One       Date:  2016-10-06       Impact factor: 3.240

4.  Effect of varying external pneumatic pressure on hemolysis and red blood cell elongation index in fresh and aged blood: Randomized laboratory research.

Authors:  Yoon Ji Choi; Hyub Huh; Go Eun Bae; Eun Ji Ko; Sung-Uk Choi; Sang-Hyun Park; Choon Hak Lim; Hye Won Shin; Hye-Won Lee; Seung Zhoo Yoon
Journal:  Medicine (Baltimore)       Date:  2018-07       Impact factor: 1.889

  4 in total