Comparison of flow rates for standard and large-bore blood tubing.

K V Iserson, A K Reeter, E Criss.   

Abstract

One of the clinically significant rate-limiting components in even the fastest of intravenous (IV) fluid systems is the tubing connecting the IV catheter to the IV solution container. We evaluated the effect large-bore tubing has on reducing this limitation on fluid flow, measuring gravity and pressurized flow in standard and large-bore tubing alone and when each was connected to an 8 French and 12-,14- and two types of 16-gauge catheters. When combined with an 8 French catheter, large-bore tubing produces a 126% increase in flow under gravity and a 90% increase under pressure. The improvement with the large tubing was less significant as the catheter size decreased. Studies in normovolemic adult volunteers achieved similar results. Large-bore blood tubing and large-bore catheters should be used in all situations where the need for rapid fluid replacement is contemplated.

Mesh:

Year:  1985        PMID: 4036117      PMCID: PMC1306267     

Source DB:  PubMed          Journal:  West J Med        ISSN: 0093-0415


  5 in total

1.  Placement of 10-gauge catheter by cutdown for rapid fluid replacement.

Authors:  J F Hansbrough; T L Cain; J S Millikan
Journal:  J Trauma       Date:  1983-03

2.  Rapid fluid resuscitation with central venous catheters.

Authors:  J R Mateer; B M Thompson; C Aprahamian; J C Darin
Journal:  Ann Emerg Med       Date:  1983-03       Impact factor: 5.721

3.  Flow rate variance of commonly used IV infusion techniques.

Authors:  D J Dula; H A Muller; J W Donovan
Journal:  J Trauma       Date:  1981-06

4.  Catheter introducers for hypovolemic patients.

Authors:  B E Haynes
Journal:  Ann Emerg Med       Date:  1982-11       Impact factor: 5.721

5.  Rapid fluid replacement: a new methodology.

Authors:  K V Iserson; A K Reeter
Journal:  Ann Emerg Med       Date:  1984-02       Impact factor: 5.721

  5 in total
  3 in total

1.  An analysis of the effect of venous resistance on the performance of gravity-fed intravenous infusion systems.

Authors:  D B Goodie; J H Philip
Journal:  J Clin Monit       Date:  1994-07

2.  Rapid fluid replacement for severe hypovolemia--human subject trials.

Authors:  K V Iserson; A K Reeter
Journal:  West J Med       Date:  1987-03

3.  Leftovers.

Authors:  David V Feliciano
Journal:  Trauma Surg Acute Care Open       Date:  2018-11-21
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.