The purpose of this study was to determine if the difference in transthoracic impedance produced by different coupling agents affects the success of shocks for defibrillation. Three different coupling agents, Harco pads (Hewlett-Packard), Littman pads (3M) and Redux paste (Hewlett-Packard), were assessed in 10 anesthetized dogs in which ventricular fibrillation was induced by electrical stimulation of the right ventricle. Defibrillation was attempted 15 seconds later, using 50, 100 and 150 joules (selected energy). Actual delivered energy, current, impedance and the percent of the shocks that achieved defibrillation were determined for the three coupling agents. Redux paste gave significantly lower impedance and higher current than the two disposable performed coupling pads tested. Despite this, there were no significant differences in shock success among the three coupling agents. Thus, in this experimental model, over a three-fold energy range, disposable coupling pads were as effective as electrode paste for defibrillation despite the slightly higher impedance of the disposable pads.

Entities:  

Mesh:

Substances:

Year:  1985        PMID: 4031281     DOI: 10.1016/s0735-1097(85)80131-5

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  5 in total

1.  Correct technique is important in cardioversion of atrial fibrillation.

Authors:  A K Slade; S M Sopher; A J Camm; D E Ward
Journal:  BMJ       Date:  1996-05-18

2.  The 1998 European Resuscitation Council guidelines for adult advanced life support. Advanced Life Support Working Group of the European Resuscitation Council.

Authors: 
Journal:  BMJ       Date:  1998-06-20

3.  Determinants of successful transthoracic defibrillation and outcome in ventricular fibrillation.

Authors:  G W Dalzell; A A Adgey
Journal:  Br Heart J       Date:  1991-06

4.  Randomised comparison of electrode positions for cardioversion of atrial fibrillation.

Authors:  T P Mathew; A Moore; M McIntyre; M T Harbinson; N P Campbell; A A Adgey; G W Dalzell
Journal:  Heart       Date:  1999-06       Impact factor: 5.994

Review 5.  A Systematic Review of the Transthoracic Impedance during Cardiac Defibrillation.

Authors:  Yasmine Heyer; Daniela Baumgartner; Christian Baumgartner
Journal:  Sensors (Basel)       Date:  2022-04-06       Impact factor: 3.576