Literature DB >> 4025750

Evaluation of the Datex 'Normac' anaesthetic agent monitor.

N P Luff, D C White.   

Abstract

The Datex 'Normac' infrared anaesthetic agent monitor has been evaluated for accuracy of calibration, stability and speed of response. Accuracy was assessed by comparing with a gas chromatograph calibrated using volumetric standards and was found to be very good for isoflurane and enflurane but was slightly less accurate for halothane, due to a low signal: noise ratio. The stability of the instrument was found to be exceptionally good. The response time is dependent on the sampling rate and we found that, in practice, the instrument would record the end expired anaesthetic concentration on patients breathing up to 15 breaths/minute. The instrument is compact and convenient for use in operating theatres.

Entities:  

Mesh:

Substances:

Year:  1985        PMID: 4025750     DOI: 10.1111/j.1365-2044.1985.tb10895.x

Source DB:  PubMed          Journal:  Anaesthesia        ISSN: 0003-2409            Impact factor:   6.955


  5 in total

1.  Methods to produce calibration mixtures for anesthetic gas monitors and how to perform volumetric calculations on anesthetic gases.

Authors:  P L Christensen; J Nielsen; T Kann
Journal:  J Clin Monit       Date:  1992-10

2.  Infrared analysis of volatile anesthetics: impact of monitor agent setting, volatile mixtures, and alcohol.

Authors:  D C Guyton; N Gravenstein
Journal:  J Clin Monit       Date:  1990-07

3.  Evaluation of three transportable multigas anesthetic monitors: the Bruel & Kjaer Anesthetic Gas Monitor 1304, the Datex Capnomac Ultima, and the Nellcor N-2500.

Authors:  J Nielsen; T Kann; J T Moller
Journal:  J Clin Monit       Date:  1993-04

4.  Accuracy and cross-sensitivity of 10 different anesthetic gas monitors.

Authors:  B Walder; R Lauber; A M Zbinden
Journal:  J Clin Monit       Date:  1993-11

5.  Continuous monitoring of alveolar and inspiratory concentrations of anesthetic and respiratory gases is difficult and potentially unsafe.

Authors:  M K Sykes
Journal:  J Clin Monit       Date:  1987-04
  5 in total