| Literature DB >> 4025176 |
Y Suzuki, H Kambara, K Kadota, S Tamaki, A Yamazato, R Nohara, G Osakada, C Kawai.
Abstract
To evaluate the noninvasive detection of shunt flow using a newly developed real-time 2-dimensional color-coded Doppler flow imaging system (D-2DE), 20 patients were examined, including 10 with secundum atrial septal defect (ASD) and 10 control subjects. These results were compared with contrast 2-dimensional echocardiography (C-2DE). Doppler 2DE displayed the blood flow toward the transducer as red and the blood flow away from the transducer as blue in 8 shades, each shade adding green according to the degree of variance in Doppler frequency. In the patients with ASD, D-2DE clearly visualized left-to-right shunt flow in 7 of 10 patients. In 5 of these 7 patients, C-2DE showed a negative contrast effect in the same area of the right atrium. Thus, D-2DE increased the sensitivity over C-2DE for detecting left-to-right shunt flow (from 50% to 70%). However, the specificity was slightly less in D-2DE (90%) than C-2DE (100%). Doppler 2DE could not visualize right-to-left shunt flow in all patients with ASD, though C-2DE showed a positive contrast effect in the left-sided heart in 9 of 10 patients with ASD. Thus, D-2DE is clinically useful for detecting left-to-right shunt flow in patients with ASD.Entities:
Mesh:
Year: 1985 PMID: 4025176 DOI: 10.1016/0002-9149(85)90862-8
Source DB: PubMed Journal: Am J Cardiol ISSN: 0002-9149 Impact factor: 2.778