Artefactual ST segment abnormalities due to electrocardiograph design.

D I Tayler, R Vincent.   

Abstract

Specifications recommended for electrocardiographs omit any reference to phase characteristics and place undue emphasis on an extended amplitude response. This does not, however, ensure faithful reproduction of the electrocardiogram and may result in less effective attenuation of unwanted noise than can be achieved with modern filters. The separate effects of phase and amplitude response on electrocardiographic signals were compared, and it was shown that distortion of the electrocardiogram by equipment with an inadequate phase response may cause changes resembling those seen in ischaemic heart disease. Case reports are presented in which distortion produced by commercial equipment impeded diagnosis. Specifications recommended as standard for electrocardiographs should be modified to include phase response and allow greater attenuation of frequencies below 0.5 Hz.

Entities:  

Mesh:

Year:  1985        PMID: 4015923      PMCID: PMC481865          DOI: 10.1136/hrt.54.2.121

Source DB:  PubMed          Journal:  Br Heart J        ISSN: 0007-0769


  10 in total

1.  Filtering without phase shift.

Authors:  R L Longini; J P Giolma; C Wall; R F Quick
Journal:  IEEE Trans Biomed Eng       Date:  1975-09       Impact factor: 4.538

2.  A spectral analysis of the normal resting electrocardiogram.

Authors:  D P Golden; R A Wolthuis; G W Hoffler
Journal:  IEEE Trans Biomed Eng       Date:  1973-09       Impact factor: 4.538

3.  A simple test for phase characteristics of chart recorders.

Authors:  J W Wagner
Journal:  IEEE Trans Biomed Eng       Date:  1979-09       Impact factor: 4.538

4.  Effect of random frequency modulation on ECG power spectra.

Authors:  J M Richardson; V K Murthy; L J Haywood
Journal:  IEEE Trans Biomed Eng       Date:  1979-02       Impact factor: 4.538

5.  Transversal resonator digital filters: fast and flexible online processors for biological signals.

Authors:  P A Lynn
Journal:  Med Biol Eng Comput       Date:  1983-11       Impact factor: 2.602

6.  Signal distortion in the electrocardiogram due to inadequate phase response.

Authors:  D Tayler; R Vincent
Journal:  IEEE Trans Biomed Eng       Date:  1983-06       Impact factor: 4.538

7.  Frequency response characteristics of ambulatory ECG monitoring systems and their implications for ST segment analysis.

Authors:  D A Bragg-Remschel; C M Anderson; R A Winkle
Journal:  Am Heart J       Date:  1982-01       Impact factor: 4.749

8.  Evaluation of autocorrelation techniques for detection of the fetal electrocardiogram.

Authors:  A G Favret; A F Caputo
Journal:  IEEE Trans Biomed Eng       Date:  1966-01       Impact factor: 4.538

9.  The low-frequency response of electrocardiographs, a frequent source of recording errors.

Authors:  A S Berson; H V Pipberger
Journal:  Am Heart J       Date:  1966-06       Impact factor: 4.749

10.  Ambulatory ST segment monitoring. Problems, pitfalls, solutions, and clinical application.

Authors:  V Balasubramanian; A Lahiri; H L Green; F D Stott; E B Raftery
Journal:  Br Heart J       Date:  1980-10
  10 in total
  6 in total

1.  Continuous ambulatory pulmonary artery pressure monitoring. A new method using a transducer tipped catheter and a simple recording system.

Authors:  R D Levy; D Cunningham; L M Shapiro; C Wright; L Mockus; K M Fox
Journal:  Br Heart J       Date:  1986-04

2.  Do we need a new standard for electrocardiographs?

Authors:  R G Gold
Journal:  Br Heart J       Date:  1985-08

3.  Effects of spinal cord stimulation on myocardial ischaemia during daily life in patients with severe coronary artery disease. A prospective ambulatory electrocardiographic study.

Authors:  M J de Jongste; J Haaksma; R W Hautvast; H L Hillege; P W Meyler; M J Staal; J E Sanderson; K I Lie
Journal:  Br Heart J       Date:  1994-05

4.  Detection of intraoperative myocardial ischaemia--a comparison among electrocardiographic, myocardial metabolic, and haemodynamic measurements in patients with reduced ventricular function.

Authors:  R I Hall; N O'Regan; M Gardner
Journal:  Can J Anaesth       Date:  1995-06       Impact factor: 5.063

5.  High-Bandpass Filters in Electrocardiography: Source of Error in the Interpretation of the ST Segment.

Authors:  F Buendía-Fuentes; M A Arnau-Vives; A Arnau-Vives; Y Jiménez-Jiménez; J Rueda-Soriano; E Zorio-Grima; A Osa-Sáez; L V Martínez-Dolz; L Almenar-Bonet; M A Palencia-Pérez
Journal:  ISRN Cardiol       Date:  2012-06-21

6.  ANMCO/AIIC/SIT Consensus Information Document: definition, precision, and suitability of electrocardiographic signals of electrocardiographs, ergometry, Holter electrocardiogram, telemetry, and bedside monitoring systems.

Authors:  Michele Massimo Gulizia; Giancarlo Casolo; Guerrino Zuin; Loredana Morichelli; Giovanni Calcagnini; Vincenzo Ventimiglia; Federica Censi; Pasquale Caldarola; Giancarmine Russo; Lorenzo Leogrande; Gian Franco Gensini
Journal:  Eur Heart J Suppl       Date:  2017-05-02       Impact factor: 1.803

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.