Literature DB >> 1152497

Failure of Beal mitral valve prosthesis. Clinical and radiographic features.

R G Jost, R C McKnight, C L Roper.   

Abstract

This report presents a case of severe dysfunction of an improved Model 104 Beall valve secondary to disc wear. This type of valve failure may be extremely difficult to detect by auscultation, echocardiography, or fluoroscopic examination. Prosthesis failure with resulting mitral insufficiency should be considered in any patient with a Teflon disc Beall mitral valve in whom there is unexplained clinical deterioration.

Entities:  

Mesh:

Year:  1975        PMID: 1152497

Source DB:  PubMed          Journal:  J Thorac Cardiovasc Surg        ISSN: 0022-5223            Impact factor:   5.209


  1 in total

1.  Mechanism of late failure of the Alvarez disc valve prosthesis.

Authors:  J L Gibbs; G A Davies; A Schofield; G A Wharton; D A Watson; L M Gerlis
Journal:  Br Heart J       Date:  1985-05
  1 in total