The effects of ionizing radiation on eight cardiac pacemakers and the influence of electromagnetic interference from two linear accelerators.

J L Venselaar.   

Abstract

Eight cardiac pacemakers were irradiated in a cobalt-60 beam. Two out of six demand-type pacemakers showed an alarming decrease in pulse repetition frequency when irradiated to dose levels that are used in radiotherapy (less than 100 Gy). Two modern programmable pacemakers showed a failure at a dose of 97 and 147 Gy, respectively. The dose levels at which these failures occurred were low enough to recommend that cardiac pacemakers should always be kept outside the radiation beam. The signals induced by electromagnetic interference (EMI) from two linear accelerators were measured using a simulation model of a pacemaker. In the laboratory, 22 modern-type pacemakers were tested with these signals to determine the sensitivity for the electromagnetic fields in the treatment rooms. It was observed that an inhibition of one pacemaker pulse was to be expected on one of the two linear accelerators when switching the machine on and off. No permanent effects were found. These findings resulted in the recommendation in our department not to use this treatment machine for radiation therapy of pacemaker-bearing patients.

Entities:  

Mesh:

Year:  1985        PMID: 3975443     DOI: 10.1016/s0167-8140(85)80011-6

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  6 in total

1.  Radiation dose monitoring in a breast cancer patient with a pacemaker: a case report.

Authors:  J R Nibhanupudy; M A de Jesus; M Fujita; A L Goldson
Journal:  J Natl Med Assoc       Date:  2001 Jul-Aug       Impact factor: 1.798

2.  Interference detection in implantable defibrillators induced by therapeutic radiation therapy.

Authors:  G J Uiterwaal; B G F Springorum; E Scheepers; G S de Ruiter; C W Hurkmans
Journal:  Neth Heart J       Date:  2006-10       Impact factor: 2.380

3.  Effects of external electrical and magnetic fields on pacemakers and defibrillators: from engineering principles to clinical practice.

Authors:  Roy Beinart; Saman Nazarian
Journal:  Circulation       Date:  2013-12-24       Impact factor: 29.690

4.  Spinal cord stimulators and radiotherapy: first case report and practice guidelines.

Authors:  Lorraine Walsh; Daipayan Guha; Thomas G Purdie; Philippe Bedard; Alexandra Easson; Fei-Fei Liu; Mojgan Hodaie
Journal:  Radiat Oncol       Date:  2011-10-25       Impact factor: 3.481

5.  Monitoring of CRT-D devices during radiation therapy in vitro.

Authors:  Martin Augustynek; David Korpas; Marek Penhaker; Jakub Cvek; Andrea Binarova
Journal:  Biomed Eng Online       Date:  2016-03-09       Impact factor: 2.819

6.  Targeted Intraoperative Radiotherapy Is a Safe Approach for Patients with Pacemakers: A Case Study and Literature Review.

Authors:  Yastira Ramdas; Carol-Ann Benn; Alexandra Grubnik; Yasmin Mayat; Dennis R Holmes
Journal:  Case Rep Oncol       Date:  2020-07-30
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.