| Literature DB >> 3964940 |
M Arakawa, L E Crooks, B McCarten, J C Hoenninger, J C Watts, L Kaufman.
Abstract
Differences in magnetic resonance (MR) imaging signal-to-noise (S/N) performance between saddle-shaped and solenoidal coils have been postulated. Each coil shape is tied to a particular magnetic field configuration, so that they are not typically interchangeable except in special situations. The solenoidal coil is predicted to have a two- to three-fold advantage over the saddle-shaped coil. Simple basic arguments raise a dispute with this assertion. Experiments show that both coils produce essentially equivalent S/N levels.Mesh:
Year: 1985 PMID: 3964940 DOI: 10.1148/radiology.154.1.3964940
Source DB: PubMed Journal: Radiology ISSN: 0033-8419 Impact factor: 11.105