The automatic implantable cardioverter-defibrillator: efficacy, complications, and device failures.

F E Marchlinski, B T Flores, A E Buxton, W C Hargrove, V P Addonizio, L W Stephenson, A H Harken, J U Doherty, E W Grogan, M E Josephson.   

Abstract

Twenty-six patients with refractory ventricular arrhythmias received the automatic implantable cardioverter-defibrillator. A patch lead only was placed during arrhythmia surgery in 7 other patients. During 13 +/- 6 (SD) months, the device discharged in 10 patients because of a sustained ventricular arrhythmia. No sudden deaths occurred. There were 31 complications in 17 patients, including postoperative refractory heart failure, coronary artery erosion, subclavian vein thrombosis, postoperative stroke after conversion of atrial fibrillation, atelectasis with pneumonia, symptomatic pleural effusions, and infection at the generator site. The cardioverter-defibrillator discharged in 9 asymptomatic patients, failed to terminate ventricular fibrillation during postoperative testing in 3 patients, and had premature battery failure in 4 patients. Tachycardia slowing during chronic amiodarone therapy and unipolar ventricular pacing during ventricular fibrillation precluded or delayed arrhythmia sensing. Thus, the cardioverter-defibrillator can be life saving, but its potential complications and interactions with antiarrhythmic drugs and pacemakers must be considered at patient selection.

Entities:  

Mesh:

Substances:

Year:  1986        PMID: 3954276     DOI: 10.7326/0003-4819-104-4-481

Source DB:  PubMed          Journal:  Ann Intern Med        ISSN: 0003-4819            Impact factor:   25.391


  8 in total

1.  Experience with unipolar pectoral defibrillation.

Authors:  R K Reddy; G H Bardy
Journal:  Herzschrittmacherther Elektrophysiol       Date:  1997-03

2.  Migration of an automatic implantable cardioverter-defibrillator patch causing massive hemothorax.

Authors:  R L Quigley; M E Hamer; S Swiryn
Journal:  Tex Heart Inst J       Date:  1996

3.  Early postoperative rise in defibrillation threshold associated with hematoma formation with unipolar defibrillation system.

Authors:  D Y Yazmajian; D Schwartzman; D J Callans; C D Gottieb; F E Marchlinski
Journal:  J Interv Card Electrophysiol       Date:  1997-09       Impact factor: 1.900

4.  Mark Josephson and the ICD: A Personal Perspective.

Authors:  Charles D Swerdlow
Journal:  Arrhythm Electrophysiol Rev       Date:  2017-04

5.  Electrode catheter ablation of resistant ventricular tachycardia in arrhythmogenic right ventricular dysplasia: experience of 15 patients with a mean follow-up of 45 months.

Authors:  G Fontaine; R Frank; I Rougier; J Tonet; Y Gallais; G Farenq; G Lascault; M Lilamand; F Fontaliran; G Chomette
Journal:  Heart Vessels       Date:  1990       Impact factor: 2.037

6.  Treatment of malignant ventricular arrhythmias with the automatic implantable cardioverter defibrillator.

Authors:  A D Slater; I Singer; C S Stavens; C Zee-Cheng; B L Ganzel; J Kupersmith; C Mavroudis; L A Gray
Journal:  Ann Surg       Date:  1989-05       Impact factor: 12.969

7.  Role of proximal electrode position in transvenous ventricular defibrillation.

Authors:  J Nitta; D S Khoury
Journal:  Ann Biomed Eng       Date:  1996 May-Jun       Impact factor: 3.934

8.  An effective technique for salvage of cardiac-related devices.

Authors:  Erin K Knepp; Karan Chopra; Hamid R Zahiri; Luther H Holton Iii; Devinder P Singh
Journal:  Eplasty       Date:  2012-01-24
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.