The use of carbon fibre material in table tops, cassette fronts and grid covers: magnitude of possible dose reduction.

A P Hufton, J G Russell.   

Abstract

The X-ray transmission of a number of radiographic components, some of conventional construction and some incorporating carbon fibre material, has been measured under clinically realistic conditions. At 80 kVp the use of carbon fibre materials enables the patient dose to be reduced by 30-50% depending on the existing equipment, type of examination and technique used. Typically the dose can be reduced by 3-15% by changing the table top, 6-12% by changing the front of the film cassette and 20-30% by using a grid with carbon fibre covers and fibre interspace. The higher cost of carbon fibre components can normally be justified by such dose savings. An indication of the absorption of all such components should be provided by manufacturers.

Entities:  

Mesh:

Substances:

Year:  1986        PMID: 3947823     DOI: 10.1259/0007-1285-59-698-157

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  3 in total

1.  Diagnostic radiography in children.

Authors:  J G Russell
Journal:  Arch Dis Child       Date:  1988-09       Impact factor: 3.791

2.  Pediatric interventional radiography equipment: safety considerations.

Authors:  Keith J Strauss
Journal:  Pediatr Radiol       Date:  2006-09

3.  Radiation dosimetry effect evaluation of a carbon fiber couch on novel uRT-linac 506c accelerator.

Authors:  Dazhen Jiang; Zhen Cao; Yongchang Wei; Tingting Cao; Jiuling Shen; Conghua Xie; Yunfeng Zhou; Hui Liu; Jun Zhang
Journal:  Sci Rep       Date:  2021-06-29       Impact factor: 4.379

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.