D P Chakraborty, E S Breatnach, M V Yester, B Soto, G T Barnes, R G Fraser.
Abstract
A modified receiver operating characteristic (ROC) study was performed in which five readers were asked to locate multiple nodules on images of an anthropomorphic phantom obtained with a prototype digital radiographic chest unit and with a conventional chest unit. Results indicate that when nodules were projected over the lungs, a significantly greater number (significant at the 5% level) were identified on conventional radiographs, whereas for nodules projected over the mediastinum, the digital images were notably superior (difference significant at the 2% level). An error analysis of the multiple nodule problem and pseudo-ROC curves are presented. The modified ROC study does not suffer from the positional ambiguity inherent in most ROC studies and is efficient in acquiring data.Mesh:
Year: 1986 PMID: 3940394 DOI: 10.1148/radiology.158.1.3940394