Literature DB >> 3831694

Chemical-shift artifact correction scheme using echo-time encoding technique.

Z H Cho, O Nalcioglu, H W Park, J B Ra, S K Hilal.   

Abstract

An investigation has been undertaken to study the effects of chemical shift on NMR image obtained by the projection reconstruction technique. Mathematical simulation studies indicate that chemical shift causes blurring in NMR images regardless of the image reconstruction technique. A systematic study of such artifacts is presented and possible correction schemes are proposed. Exemplary results with and without correction of the chemical-shift artifact and pulse schemes employed for projection reconstruction are presented. The present chemical-shift correction scheme is basically the same as a spectroscopic imaging technique such as the existing 4-D NMR imaging method except that the encoding steps required can be reduced substantially in the case of modest spectroscopic resolution.

Mesh:

Year:  1985        PMID: 3831694     DOI: 10.1002/mrm.1910020308

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  2 in total

1.  1H-spectroscopic imaging with read gradient during acquisition in inhomogeneous fields: analysis, measurement strategy, and data processing.

Authors:  J Weis; A Ericsson; A Hemmingsson
Journal:  MAGMA       Date:  1997-09       Impact factor: 2.310

2.  High spectral and spatial resolution MRI of age-related changes in murine prostate.

Authors:  Sean Foxley; Xiaobing Fan; Sanaz A Jansen; Marta Zamora; Erica Markiewicz; Hikmat Al-Ahmadie; Gregory S Karczmar
Journal:  Magn Reson Med       Date:  2008-09       Impact factor: 4.668

  2 in total