Continuous in vivo oxygen saturation in newborn infants with pulmonary disease: a new fiberoptic catheter oximeter.

A R Wilkinson, R H Phibbs, G A Gregory.   

Abstract

We measured the in vivo oxygen saturation of hemoglobin in the arterial blood of 34 sick newborn infants with a new, rapidly responding, continuously recording, catheter oximeter. The oxygen saturation, SaO2, was found to fluctuate between 85 and 95% in infants with pulmonary disease who are in a stable condition and breathing spontaneously. Severe disaturation occurred during apnea and after procedures such as chest percussion and suctioning of the endotrachael tube. The effects of changes in respiratory therapy were quickly manifested by changes in SaO2. This was particularly useful in guiding resuscitation of newborn infants who were asphyxiated at birth. Blood sampling for measurement of PaO2 can be done less frequently, reducing the need for blood transfusions in small infants.

Entities:  

Mesh:

Substances:

Year:  1979        PMID: 380903     DOI: 10.1097/00003246-197905000-00005

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  4 in total

1.  Continuous monitoring of mixed venous oxygen saturation in infants after cardiac surgery.

Authors:  D Schranz; S Schmitt; H Oelert; F Schmid; R Huth; B Zimmer; A Schuind; K Vogel; H Stopfkuchen; B K Jüngst
Journal:  Intensive Care Med       Date:  1989       Impact factor: 17.440

2.  Continuous central venous and pulmonary artery oxygen saturation monitoring in the critically ill.

Authors:  Konrad Reinhart; Hans-Jörg Kuhn; Christiane Hartog; Donald L Bredle
Journal:  Intensive Care Med       Date:  2004-06-09       Impact factor: 17.440

3.  Pulse oximeter and transcutaneous arterial oxygen measurements in neonatal and paediatric intensive care.

Authors:  D P Southall; S Bignall; V A Stebbens; J R Alexander; R P Rivers; T Lissauer
Journal:  Arch Dis Child       Date:  1987-09       Impact factor: 3.791

Review 4.  Continuous mixed venous oxygen saturation measurement: a significant advance in hemodynamic monitoring?

Authors:  E A Norfleet; C B Watson
Journal:  J Clin Monit       Date:  1985-10
  4 in total