Literature DB >> 3691403

Color performance of video endoscopes: quantitative measurement of color reproduction.

K Knyrim1, H K Seidlitz, F Hagenmüller, M Classen.   

Abstract

We have conducted a study aimed at testing the color performance of video endoscopes in quantitative terms for the first time. The video endoscopes investigated were from Fuji, Olympus, Toshiba/Machida, and Welch-Allyn. The tests were carried out with an opto-electronic color analyser using standardized color charts and a neutral grey chart. Hue and chroma of the reproduced colors were measured. The Olympus video endoscope reproduces the hue very well, showing only slight color desaturation. The Fuji video endoscope had a yellowish tint, whereas the picture obtained with the WA video endoscope produced an inhomogeneous picture with a pronounced red/purplish cast at the top of the screen and almost neutral reproduction at the lower half. The Toshiba video endoscope was the only one with a color chip, and reproduced hue fairly well, but its colors were weakly saturated.

Mesh:

Year:  1987        PMID: 3691403     DOI: 10.1055/s-2007-1018290

Source DB:  PubMed          Journal:  Endoscopy        ISSN: 0013-726X            Impact factor:   10.093


  2 in total

1.  Colonoscope flexural rigidity measurement.

Authors:  J A Wehrmeyer; J A Barthel; J P Roth; T Saifuddin
Journal:  Med Biol Eng Comput       Date:  1998-07       Impact factor: 2.602

2.  Electronic imaging in endoscopy.

Authors:  G Berci; M Paz-Partlow
Journal:  Surg Endosc       Date:  1988       Impact factor: 4.584

  2 in total