Tony L Buhr1, Erica Borgers-Klonkowski1, Bradford W Gutting1, Emlyn E Hammer1, Shelia M Hamilton1, Brett M Huhman2, Stuart L Jackson2, Neil L Kennihan1, Samuel D Lilly1, John D Little2, Brooke B Luck1, Emily A Matuczinski1, Charles T Miller1, Rachel E Sides1, Vanessa L Yates1, Alice A Young1.
Abstract
Aims: The dosages and efficacy of 14 ultraviolet (UV) decontamination technologies were measured against a SARS-CoV-2 surrogate virus that was dried onto different materials for laboratory and field testing. Methods and results: A live enveloped, ribonucleic acid (RNA) virus surrogate for SARS-CoV-2 was dried on stainless steel 304 (SS304), Navy Top Coat-painted SS304 (NTC), cardboard, polyurethane, polymethyl methacrylate (PMMA), and acrylonitrile butadiene styrene (ABS) materials at > 8.0 log10 plaque-forming units (PFU) per test coupon. The coupons were then exposed to UV radiation during both laboratory and field testing. Commercial and prototype UV-emitting devices were measured for efficacy: four handheld devices, three room/surface-disinfecting machines, five air disinfection devices, and two larger custom-made machines. UV device dosages ranged from 0.01 to 729 mJ cm-2. The antiviral efficacy among the different UV devices ranged from no decontamination up to nearly achieving sterilization. Importantly, cardboard required far greater dosage than SS304.Entities:
Keywords: COVID-19; SARS-CoV-2; UV decontamination; decontamination; enveloped virus; enveloped virus Ф6; ultraviolet (UV)
Year: 2022 PMID: 36267449 PMCID: PMC9578676 DOI: 10.3389/fbioe.2022.875817
Source DB: PubMed Journal: Front Bioeng Biotechnol ISSN: 2296-4185
FIGURE 1Φ6 extraction procedure for test samples and non-treated controls.
FIGURE 2Normalized wavelength emission spectra for 14 UV devices. Top frame shows Xenex ( ) and pulsed xenon ( ) emission. Lower frame shows the emission for three 222 nm emitters ( ), eight 254 nm emitters ( ), and the 272 nm emitter ( ).
Average extraction efficiency of Φ6 virus 4–14 days after drying on different materials.
| PMMA | ABS | ||||||
|---|---|---|---|---|---|---|---|
| Plastic | Plastic | Cardboard | NTC | SS304 | Polyurethane | Glass | |
| Test days | 11 | 4 | 16 | 15 | 18 | 9 | 1 |
| Total samples | 55 | 20 | 80 | 75 | 90 | 45 | 5 |
| Extraction % | 57.1 ± 24.2 | 71.8 ± 14.6 | 59.3 ± 18.7 | 54.3 ± 20.0 | 58.1 ± 19.9 | 54.1 ± 18.8 | 95.2 ± 10.2 |
FIGURE 3Testing setup and UV coverage for 18 and 35 W handheld devices. (A) GermAwayUV 18 W handheld device and custom test chamber shown during coupon exposure. (B) GermAwayUV 35 W handheld device and custom test chamber, shown in the pre/post exposure state. (C,D) UV coverage heat maps for the 18 W device (C) and the 35 W device (D) taken 5 cm from the source.
Dosage and efficacy of handheld, room, and chamber-type devices showing log10 reduction data based on the steady-state emission, not peak emission. White (low decontamination) = Fail <2 log10; yellow (sanitation) = Fail ≥2 log10, <3 log10; light blue (disinfection) = Pass ≥3 log10; and dark blue (approaching virus sterilization) = Pass ≥6 log10. N/A—dosage measurements had no meaning because of the broad-spectrum Xe source.
| Name | Description | Dosage (mJ cm−2) | Exposure distance | Exposure time | Efficacy (log10 reduction after an >8 log10 challenge of live Ф6) | ||||
|---|---|---|---|---|---|---|---|---|---|
| SS304 | NTC | Cardboard |
| Polyurethane | |||||
| GermAwayUV 18 W handheld | Two 12.7 cm Hg U-shape bulbs in the handheld device 254 nm UV-C MSRP $100 | 101.2 | 5 cm | 10 s | 2.5 ± 0.1 | 2.0 ± 0.0 | 1.9 ± 0.0 | 1.9 ± 0.0 | 2.0 ± 0.0 |
| 202.4 | 5 cm | 20 s | 4.3 ± 0.2 | 3.1 ± 0.2 | 2.0 ± 0.1 | 4.5 ± 0.1 | 3.3 ± 0.1 | ||
| GermAway UV 35 W handheld | Two 22.5 cm Hg twin tube bulbs in the handheld device 254 nm UV-C MSRP $450 | 13.8 | 5 cm | 2 s | 0.3 ± 0.0 | 0.2 ± 0.0 | 0.3 ± 0.0 | 0.3 ± 0.0 | 0.3 ± 0.0 |
| 34.5 | 5 cm | 5 s | 0.9 ± 0.1 | 0.6 ± 0.0 | 0.7 ± 0.0 | 0.8 ± 0.1 | 0.7 ± 0.0 | ||
| 69.0 | 5 cm | 10 s | 1.4 ± 0.1 | 1.2 ± 0.1 | 1.2 ± 0.0 | 1.3 ± 0.0 | 1.1 ± 0.1 | ||
| 272 nm prototype handheld | Eight LED strips divided by angled plastic in the handheld device 272 nm UV-C | 31.2 | 5 cm | 2 s | 3.0 ± 0.3 | 1.6 ± 0.2 | 1.6 ± 0.1 | 1.6 ± 0.2 | 1.6 ± 0.2 |
| 78.0 | 5 cm | 5 s | 5.2 ± 0.1 | 2.8 ± 0.1 | 2.5 ± 0.2 | 3.1 ± 0.2 | 2.0 ± 0.9 | ||
| 156 | 5 cm | 10 s | 6.2 ± 0.4 | 3.8 ± 0.2 | 2.4 ± 0.1 | 4.9 ± 0.2 | 4.7 ± 0.5 | ||
| 222 nm excimer prototype handheld | Three lamp modules attached to 2.54 cm thick plastic panel, in the handheld device 222 nm UV-C | 5.9 | 5 cm | 2 s | 0.6 ± 0.1 | 0.4 ± 0.2 | 0.1 ± 0.1 | 0.3 ± 0.1 | 0.3 ± 0.1 |
| 14.8 | 5 cm | 5 s | 0.9 ± 0.2 | 0.1 ± 0.1 | 0.5 ± 0.1 | 0.5 ± 0.1 | 0.6 ± 0.2 | ||
| 29.6 | 5 cm | 10 s | 1.1 ± 0.2 | 0.9 ± 0.1 | 0.6 ± 0.1 | 0.9 ± 0.1 | 1.1 ± 0.1 | ||
| Mounted pulsed-xenon prototype | Pulsed Xe bulb in small housing ceiling, wall, or tripod-mounted broad-spectrum UV-B and UV-C | N/A | 0.5 m | 15 min | 1.4 ± 0.1 | 1.1 ± 0.1 | 1.1 ± 0.1 | 1.5 ± 0.1 | N/A |
| 0.5 m | 30 min | 2.3 ± 0.1 | 2.3 ± 0.1 | 1.6 ± 0.1 | 2.9 ± 0.1 | N/A | |||
| 0.5 m | 60 min | 5.2 ± 0.2 | 3.2 ± 0.3 | 2.4 ± 0.2 | 5.2 ± 0.2 | N/A | |||
| 1 m | 15 min | 0.5 ± 0.0 | 0.4 ± 0.1 | 0.4 ± 0.1 | 0.3 ± 0.1 | N/A | |||
| 1 m | 30 min | 0.8 ± 0.0 | 0.7 ± 0.0 | 0.7 ± 0.1 | 0.7 ± 0.0 | N/A | |||
| 1 m | 60 min | 1.5 ± 0.1 | 1.3 ± 0.1 | 1.0 ± 0.0 | 1.3 ± 0.1 | N/A | |||
| 2 m | 15 min | 0.0 ± 0.1 | 0.1 ± 0.1 | 0.2 ± 0.0 | 0.1 ± 0.0 | N/A | |||
| 2 m | 30 min | 0.1 ± 0.0 | 0.0 ± 0.0 | 0.2 ± 0.0 | 0.2 ± 0.1 | N/A | |||
| 2 m | 60 min | 0.3 ± 0.0 | 0.2 ± 0.1 | 0.3 ± 0.0 | 0.2 ± 0.0 | N/A | |||
| Xenex LightStrike | One pulsed Xe bulb mounted on the rolling cart broad-spectrum UV-B and UV-C MSRP $125,000 | N/A | 178 cm | 5 min | 0.8 ± 0.1 | 0.4 ± 0.0 | 0.5 ± 0.0 | 0.5 ± 0.0 | N/A |
| N/A | 178 cm | 20 min | 1.7 ± 0.0 | 1.5 ± 0.1 | 1.3 ± 0.0 | 2.2 ± 0.1 | NA | ||
| Light emitting module (LEM) | 20 Hg bulbs mounted in a ring on the rolling cart 254 nm UV-C MSRP $95,000 | 60 | 263 cm | 4 min 22 s | 3.1 ± 0.3 | 2.6 ± 0.2 | 1.6 ± 0.1 | 2.5 ± 0.2 | 2.1 ± 0.1 |
| 100 | 263 cm | 7 min 2 s | 4.6 ± 0.5 | 2.7 ± 0.2 | 2.2 ± 0.1 | 2.7 ± 0.2 | 3.1 ± 0.1 | ||
| 140 | 263 cm | 9 min 33 s | 5.3 ± 0.5 | 3.4 ± 0.1 | 2.7 ± 0.1 | 4.0 ± 0.2 | 3.9 ± 0.1 | ||
| Medium conveyer prototype | Chamber lined on four sides with Hg bulbs and powered conveyer belt to move items through 254 nm UV-C | 60 | 62 cm | 20 s | 6.1 ± 0.2 | 4.5 ± 0.1 | 2.9 ± 0.3 | 4.8 ± 0.2 | 5.3 ± 0.1 |
| 100 | 62 cm | 32 s | 7.6 ± 0.3 | 4.4 ± 0.2 | 2.9 ± 0.2 | 6.3 ± 0.2 | 7.1 ± 0.3 | ||
| 140 | 62 cm | 44 s | 7.1 ± 0.4 | 4.3 ± 0.5 | 2.6 ± 0.1 | 6.3 ± 0.3 | 6.3 ± 0.3 | ||
| 13.7 | 62 cm | 8 s | 1.6 ± 0.2 | 1.4 ± 0.1 | 1.3 ± 0.1 | 1.4 ± 0.1 | N/A | ||
| 23.8 | 62 cm | 16 s | 2.7 ± 0.2 | 1.9 ± 0.1 | 1.8 ± 0.1 | 2.1 ± 0.1 | N/A | ||
| 40.0 | 62 cm | 24 s | 3.5 ± 0.3 | 2.8 ± 0.1 | 2.5 ± 0.1 | 2.9 ± 0.1 | N/A | ||
| 56.2 | 62 cm | 32 s | 4.7 ± 0.4 | 3.2 ± 0.2 | 2.9 ± 0.1 | 3.7 ± 0.1 | N/A | ||
| Big box prototype | Chamber lined on sides and top with Hg bulbs (total 320) 254 nm UV-C | 377–729 | 17 cm | 2 min | 5.4 ± 0.2 | 3.6 ± 0.3 | 3.1 ± 0.2 | 5.7 ± 0.1 | 5.4 ± 0.3 |
ABS plastic was tested for the mounted pulsed Xe and Big Box prototypes. PMMA plastic was tested for the other devices.
Dosage and efficacy of handheld, room, and chamber-type devices showing log10 survival data based on the steady state emission, not peak emission. White (low decontamination) = Fail <2 log10; yellow (sanitation) = Fail ≥2 log10, <3 log10; light blue (disinfection) = Pass ≥3 log10; and dark blue (approaching virus sterilization) = Pass ≥6 log10. N/A—dosage measurements had no meaning because of the broad-spectrum Xe source.
| Name | Description | Dosage (mJ cm−2) | Exposure distance | Exposure time | Log10 survival after a >8 log10 challenge of live Ф6 | |||||
|---|---|---|---|---|---|---|---|---|---|---|
| Inoculum | SS304 | NTC | Cardboard |
| Polyurethane | |||||
| GermAwayUV 18 W handheld | Two 12.7 cm U-shape Hg bulbs in the handheld device 254 nm UV-C MSRP $100 | 101.2 | 5 cm | 10 s | 8.4 ± 0.1 | 5.9 ± 0.5 | 6.4 ± 0.1 | 6.5 ± 0.1 | 5.5 ± 1.1 | 6.4 ± 0.1 |
| 202.4 | 5 cm | 20 s | 8.2 ± 0.0 | 3.9 ± 0.4 | 5.0 ± 0.4 | 6.1 ± 0.3 | 3.7 ± 0.3 | 5.1 ± 0.2 | ||
| GermAway UV 35 W handheld | Two 22.5 cm twin tube Hg bulbs in the handheld device 254 nm UV-C MSRP $450 | 13.8 | 5 cm | 2 s | 8.2 ± 0.1 | 7.8 ± 0.2 | 7.9 ± 0.1 | 7.8 ± 0.1 | 7.8 ± 0.2 | 7.8 ± 0.1 |
| 34.5 | 5 cm | 5 s | 7.2 ± 0.2 | 7.5 ± 0.1 | 7.3 ± 0.1 | 7.3 ± 0.3 | 7.4 ± 0.1 | |||
| 69.0 | 5 cm | 10 s | 6.6 ± 0.3 | 6.9 ± 0.2 | 6.9 ± 0.1 | 6.8 ± 0.2 | 7.0 ± 0.2 | |||
| 272 nm prototype handheld | Eight LED strips divided by angled plastic in the handheld device 272 nm UV-C | 31.2 | 5 cm | 2 s | 8.5 ± 0.2 | 5.5 ± 0.6 | 7.0 ± 0.3 | 6.9 ± 0.2 | 6.9 ± 0.3 | 6.9 ± 0.4 |
| 78.0 | 5 cm | 5 s | 3.3 ± 0.0 | 5.7 ± 0.1 | 6.0 ± 0.4 | 5.4 ± 0.3 | 6.5 ± 1.9 | |||
| 156 | 5 cm | 10 s | 2.3 ± 0.7 | 4.7 ± 0.3 | 6.1 ± 0.1 | 3.6 ± 0.5 | 3.8 ± 0.9 | |||
| 222 nm excimer prototype handheld | Three lamp modules attached to 2.54 cm thick plastic panel, in the handheld device 222 nm UV-C | 5.9 | 5 cm | 2 s | 8.3 ± 0.2 | 7.7 ± 0.2 | 7.9 ± 0.3 | 8.2 ± 0.1 | 8.0 ± 0.1 | 7.9 ± 0.0 |
| 14.8 | 5 cm | 5 s | 7.4 ± 0.3 | 8.2 ± 0.1 | 7.8 ± 0.1 | 7.8 ± 0.2 | 7.6 ± 0.3 | |||
| 29.6 | 5 cm | 10 s | 7.1 ± 0.4 | 7.1 ± 0.4 | 7.6 ± 0.1 | 7.4 ± 0.2 | 7.1 ± 0.1 | |||
| Mounted pulsed xenon prototype | Pulsed Xe bulb in small housing ceiling, wall, or tripod-mounted broad-spectrum UV-B and UV-C | N/A | 0.5 m | 15 min | 8.2 ± 0.1 | 6.8 ± 0.2 | 7.0 ± 0.2 | 7.1 ± 0.2 | 6.7 ± 0.1 | N/A |
| 0.5 m | 30 min | 5.9 ± 0.3 | 5.8 ± 0.3 | 6.6 ± 0.1 | 5.3 ± 0.2 | N/A | ||||
| 0.5 m | 60 min | 2.9 ± 0.4 | 5.0 ± 0.8 | 5.7 ± 0.5 | 3.0 ± 0.6 | N/A | ||||
| 1 m | 15 min | 7.7 ± 0.1 | 7.7 ± 0.2 | 7.7 ± 0.1 | 7.8 ± 0.1 | N/A | ||||
| 1 m | 30 min | 7.4 ± 0.1 | 7.5 ± 0.0 | 7.4 ± 0.1 | 7.4 ± 0.1 | N/A | ||||
| 1 m | 60 min | 6.7 ± 0.2 | 6.8 ± 0.2 | 7.2 ± 0.1 | 6.8 ± 0.2 | N/A | ||||
| 2 m | 15 min | 8.2 ± 0.1 | 8.1 ± 0.1 | 8.0 ± 0.1 | 8.1 ± 0.0 | N/A | ||||
| 2 m | 30 min | 8.1 ± 0.1 | 8.1 ± 0.1 | 8.0 ± 0.1 | 8.0 ± 0.1 | N/A | ||||
| 2 m | 60 min | 7.9 ± 0.0 | 7.9 ± 0.1 | 7.8 ± 0.1 | 7.9 ± 0.1 | N/A | ||||
| Xenex LightStrike | One pulsed-Xe bulb mounted on the rolling cart broad-spectrum UV-B and UV-C MSRP $125,000 | N/A | 178 cm | 5 min | 8.4 ± 0.0 | 7.6 ± 0.2 | 8.0 ± 0.1 | 7.9 ± 0.1 | 7.8 ± 0.1 | N/A |
| N/A | 178 cm | 20 min | 6.7 ± 0 | 6.9 ± 0.3 | 7.1 ± 0.0 | 6.2 ± 0.1 | N/A | |||
| Light emitting module (LEM) | 20 Hg bulbs mounted in a ring on the rolling cart 254 nm UV-C MSRP $95,000 | 60 | 263 cm | 4 min 22 s | 8.4 ± 0.2 | 5.3 ± 0.6 | 5.8 ± 0.5 | 6.8 ± 0.2 | 5.9 ± 0.5 | 6.3 ± 0.1 |
| 100 | 263 cm | 7 min 2 s | 3.8 ± 1.1 | 5.7 ± 0.5 | 6.2 ± 0.2 | 5.7 ± 0.4 | 5.3 ± 0.1 | |||
| 140 | 263 cm | 9 min 33 s | 3.1 ± 1.1 | 5.0 ± 0.3 | 5.8 ± 0.2 | 4.4 ± 0.3 | 4.6 ± 0.2 | |||
| Medium conveyer prototype | Chamber lined on four sides with Hg bulbs and powered conveyer belt to move items through 254 nm UV-C | 60 | 62 cm | 20 s | 8.2 ± 0.1 | 2.1 ± 0.4 | 3.7 ± 0 | 5.3 ± 0.6 | 3.4 ± 0.4 | 2.9 ± 0.1 |
| 100 | 62 cm | 32 s | 0.6 ± 0.6 | 3.8 ± 0.5 | 5.3 ± 0.4 | 2.0 ± 0.4 | 1.1 ± 0.7 | |||
| 140 | 62 cm | 44 s | 1.1 ± 0.8 | 3.9 ± 1.1 | 5.6 ± 0.1 | 1.9 ± 0.6 | 0.9 ± 0.5 | |||
| 13.7 | 62 cm | 8 s | 8.4 ± 0.2 | 6.8 ± 0.4 | 7.0 ± 0.1 | 7.1 ± 0.1 | 7.0 ± 0.1 | N/A | ||
| 23.8 | 62 cm | 16 s | 5.6 ± 0.4 | 6.5 ± 0.1 | 6.5 ± 0.2 | 6.3 ± 0.1 | N/A | |||
| 40.0 | 62 cm | 24 s | 4.8 ± 0.6 | 5.5 ± 0.2 | 5.8 ± 0.3 | 5.4 ± 0.2 | N/A | |||
| 56.2 | 62 cm | 32 s | 3.7 ± 1.0 | 5.1 ± 0.4 | 5.5 ± 0.2 | 4.7 ± 0.1 | N/A | |||
| Big box prototype | Chamber lined on sides and top with Hg bulbs (total 320) 254 nm UV-C | 377–729 | 17 cm | 2 min | 8.4 ± 0.1 | 3.0 ± 0.4 | 4.8 ± 0.6 | 5.3 ± 0.5 | 2.5 ± 0.3 | 3.0 ± 0.6 |
| 8.2 ± 0.1 (ABS plastic only) | ||||||||||
ABS plastic was tested for the mounted pulsed Xe and the Big Box prototypes. PMMA plastic was tested for the other devices.
FIGURE 4(A) Prototype 272 nm LED handheld inside the wooden test chamber. (B) UV coverage heat map taken 5 cm from the source.
FIGURE 5(A) Prototype 222 nm Excimer Lamp Module Board inside the wooden test chamber. (B) UV coverage heat map taken 5 cm from the source.
FIGURE 6Virus-inoculated coupons were placed 0.5, 1, or 2 m below the pulsed Xe prototype.
FIGURE 7Virus-inoculated coupons were placed 1.78 m horizontally from the Xenex LightStrike or 2.63 m away from the LEM.
FIGURE 8Dosimeters traveling down conveyor to determine exposure times for target UV dosages.
FIGURE 9Prototype Big Box UV-C Chamber with a double-stacked pallet (top). Coupon placement was inside Petri plates on top of the pallet for testing (bottom).
Dosage and efficacy of room air-irradiating prototypes showing log10 reduction data based on the average power from the bulb area, not the peaks. White (low decontamination) = Fail <2 log10; yellow (sanitation) = Fail ≥2 log10, <3 log10; light blue (disinfection) = Pass ≥3 log10; and dark blue (approaching virus sterilization) = Pass ≥6 log10. N/A—dosage measurements had no meaning because of the broad-spectrum Xe source.
| Name | Description | Dosage (mJ cm−2) | Exposure distance (cm) | Exposure time (s) | Efficacy (log10 reduction after a >8 log10 challenge of live Ф6) | |
|---|---|---|---|---|---|---|
| SS 304 | Quartz glass | |||||
| Ceiling mounted prototype A: original | Two Hg bulbs within an enclosed chamber and fans circulate air through unit 254 nm UV-C | 18.6 | 5 | 5 | 0.9 ± 0.1 | NA |
| 37.2 | 10 | 1.3 ± 0.1 | NA | |||
| 55.8 | 15 | 1.7 ± 0.1 | NA | |||
| 12.9 | 10 | 5 | 0.6 ± 0.1 | NA | ||
| 25.7 | 10 | 1.0 ± 0.1 | NA | |||
| 38.6 | 15 | 1.3 ± 0.1 | NA | |||
| 10.6 | 15 | 5 | 0.5 ± 0.1 | NA | ||
| 21.1 | 10 | 0.8 ± 0.1 | NA | |||
| 31.7 | 15 | 1.1 ± 0.1 | NA | |||
| Wall-mounted device A: modified | KrCl/excimer lamp within enclosed chamber and fans circulate air through unit 222 nm UV-C | 15.2 | 4 | 5 | NA | 5.5 ± 0.6 |
| 30.5 | 10 | NA | 6.4 ± 0.3 | |||
| 45.8 | 15 | NA | 6.6 ± 0.7 | |||
| 6.9 | 10 | 5 | NA | 2.2 ± 0.2 | ||
| 13.7 | 10 | NA | 4.6 ± 0.7 | |||
| 20.6 | 15 | NA | 6.7 ± 0.2 | |||
| 3.4 | 20 | 5 | NA | 1.3 ± 0.1 | ||
| 6.8 | 10 | NA | 2.4 ± 0.3 | |||
| 10.2 | 15 | NA | 4.0 ± 0.6 | |||
| Wall-mounted prototype B: original | Hg bulb within open sconce and fans circulate air through unit 254 nm UV-C | 29.9 | 5 | 5 | 1.6 ± 0.1 | NA |
| 59.8 | 10 | 2.5 ± 0.1 | NA | |||
| 89.7 | 15 | 3.4 ± 0.1 | NA | |||
| 15.8 | 10 | 5 | 1.1 ± 0.1 | NA | ||
| 31.6 | 10 | 1.7 ± 0.1 | NA | |||
| 47.4 | 15 | 2.3 ± 0.1 | NA | |||
| 9.9 | 15 | 5 | 0.8 ± 0.1 | NA | ||
| 19.7 | 10 | 1.3 ± 0.2 | NA | |||
| 29.6 | 15 | 1.5 ± 0.1 | NA | |||
| Wall-mounted prototype B: modified | KrCl/excimer lamp within open sconce and fans circulate air through unit 222 nm UV-C | 15.6 | 5 | 5 | 0.5 ± 0.1 | NA |
| 31.2 | 10 | 0.9 ± 0.2 | NA | |||
| 46.8 | 15 | 1.2 ± 0.3 | NA | |||
| 7.5 | 15 | 5 | 0.6 ± 0.2 | NA | ||
| 15.0 | 10 | 0.5 ± 0.3 | NA | |||
| 22.5 | 15 | 0.9 ± 0.1 | NA | |||
| 2.8 | 30.5 | 5 | 0.1 ± 0.1 | NA | ||
| 5.6 | 10 | 0.2 ± 0.1 | NA | |||
| 8.4 | 15 | 0.2 ± 0.2 | NA | |||
| 0.2 | 61 | 5 | 0.0 ± 0.2 | NA | ||
| 0.3 | 10 | 0.0 ± 0.2 | NA | |||
| 0.5 | 15 | 0.1 ± 0.1 | NA | |||
| 0.1 | 71 | 5 | 0.0 ± 0.1 | NA | ||
| 0.2 | 10 | 0.0 ± 0.1 | NA | |||
| 0.3 | 15 | 0.0 ± 0.1 | NA | |||
| Wall-mounted prototype C | Hg bulb within open sconce and fans circulate air through unit 254 nm UV-C | 16.6 | 5 | 5 | 0.9 ± 0.2 | NA |
| 33.2 | 10 | 1.6 ± 0.0 | NA | |||
| 49.8 | 15 | 2.2 ± 0.2 | NA | |||
| 9.1 | 10 | 5 | 0.7 ± 0.1 | NA | ||
| 18.2 | 10 | 0.9 ± 0.1 | NA | |||
| 27.3 | 15 | 1.3 ± 0.1 | NA | |||
| 6.4 | 15 | 5 | 0.5 ± 0.1 | NA | ||
| 12.8 | 10 | 0.8 ± 0.0 | NA | |||
| 19.2 | 15 | 1.2 ± 0.1 | NA | |||
Dosage and efficacy of room air-irradiating prototypes showing log10 survival data based on the average power from the bulb area, not the peaks. White (low decontamination) = Fail <2 log10; yellow (sanitation) = Fail ≥2 log10, <3 log10; light blue (disinfection) = Pass ≥3 log10; and dark blue (approaching virus sterilization) = Pass ≥6 log10. N/A—dosage measurements had no meaning because of the broad-spectrum Xe source.
| Name | Description | Dosage (mJ cm−2) | Exposure distance (cm) | Exposure time (s) | Log10 survival after a >8 log10 challenge of live Ф6 | ||
|---|---|---|---|---|---|---|---|
| Inoculum | SS304 | Quartz glass | |||||
| Ceiling-mounted prototype A: original | Two Hg bulbs within enclosed chamber and fans circulate air through unit 254 nm UV-C | 18.6 | 5 | 5 | 8.2 ± 0.0 | 7.3 ± 0.3 | NA |
| 37.2 | 10 | 6.9 ± 0.1 | NA | ||||
| 55.8 | 15 | 6.5 ± 0.2 | NA | ||||
| 12.9 | 10 | 5 | 7.6 ± 0.2 | NA | |||
| 25.7 | 10 | 7.2 ± 0.3 | NA | ||||
| 38.6 | 15 | 7.0 ± 0.1 | NA | ||||
| 10.6 | 15 | 5 | 7.8 ± 0.1 | NA | |||
| 21.1 | 10 | 7.4 ± 0.3 | NA | ||||
| 31.7 | 15 | 7.1 ± 0.2 | NA | ||||
| Wall-mounted prototype A: modified | KrCl/excimer lamp within enclosed chamber and fans circulate air through unit 222 nm UV-C | 15.2 | 4 | 5 | 8.2 ± 0.1 | NA | 2.7 ± 1.4 |
| 30.5 | 10 | NA | 1.8 ± 0.7 | ||||
| 45.8 | 15 | NA | 1.6 ± 1.6 | ||||
| 6.9 | 10 | 5 | NA | 6.0 ± 0.5 | |||
| 13.7 | 10 | NA | 3.5 ± 1.5 | ||||
| 20.6 | 15 | NA | 1.4 ± 0.4 | ||||
| 3.4 | 20 | 5 | NA | 6.8 ± 0.3 | |||
| 6.8 | 10 | NA | 5.7 ± 0.7 | ||||
| 10.2 | 15 | NA | 4.1 ± 1.3 | ||||
| Wall-mounted prototype B: original | Hg bulb within open sconce and fans circulate air through unit 254 nm UV-C | 29.9 | 5 | 5 | 8.2 ± 0.0 | 6.7 ± 0.1 | NA |
| 59.8 | 10 | 5.7 ± 0.2 | NA | ||||
| 89.7 | 15 | 4.8 ± 0.3 | NA | ||||
| 15.8 | 10 | 5 | 7.1 ± 0.2 | NA | |||
| 31.6 | 10 | 6.5 ± 0.2 | NA | ||||
| 47.4 | 15 | 5.9 ± 0.3 | NA | ||||
| 9.9 | 15 | 5 | 7.5 ± 0.2 | NA | |||
| 19.7 | 10 | 6.9 ± 0.4 | NA | ||||
| 29.6 | 15 | 6.7 ± 0.2 | NA | ||||
| Wall-mounted prototype B: modified | KrCl/excimer lamp within open sconce and fans circulate air through unit 222 nm UV-C | 15.6 | 5 | 5 | 8.1 ± 0.3 | 7.6 ± 0.1 | NA |
| 31.2 | 10 | 7.2 ± 0.4 | NA | ||||
| 46.8 | 15 | 6.9 ± 0.6 | NA | ||||
| 7.5 | 15 | 5 | 7.6 ± 0.2 | NA | |||
| 15.0 | 10 | 7.6 ± 0.6 | NA | ||||
| 22.5 | 15 | 7.2 ± 0.1 | NA | ||||
| 2.8 | 30.5 | 5 | 8.1 ± 0.1 | NA | |||
| 5.6 | 10 | 7.9 ± 0.2 | NA | ||||
| 8.4 | 15 | 7.9 ± 0.2 | NA | ||||
| 0.2 | 61 | 5 | 8.1 ± 0.2 | NA | |||
| 0.3 | 10 | 8.1 ± 0.2 | NA | ||||
| 0.5 | 15 | 8.0 ± 0.1 | NA | ||||
| 0.1 | 71 | 5 | 8.1 ± 0.1 | NA | |||
| 0.2 | 10 | 8.1 ± 0.1 | NA | ||||
| 0.3 | 15 | 8.1 ± 0.2 | NA | ||||
| Wall-mounted prototype C | Hg bulb within open sconce and fans circulate air through unit 254 nm UV-C | 16.6 | 5 | 5 | 8.2 ± 0.0 | 7.4 ± 0.3 | NA |
| 33.2 | 10 | 6.7 ± 0.1 | NA | ||||
| 49.8 | 15 | 6.1 ± 0.3 | NA | ||||
| 9.1 | 10 | 5 | 7.5 ± 0.2 | NA | |||
| 18.2 | 10 | 7.3 ± 0.2 | NA | ||||
| 27.3 | 15 | 6.9 ± 0.2 | NA | ||||
| 6.4 | 15 | 5 | 7.7 ± 0.1 | NA | |||
| 12.8 | 10 | 7.4 ± 0.1 | NA | ||||
| 19.2 | 15 | 7.1 ± 0.1 | NA | ||||