Bhim B Rai1,2, Faran Sabeti1,3, Corinne F Carle1, Emilie M Rohan1, Josh P van Kleef1, Rohan W Essex4, Richard C Barry5, Ted Maddess1.
Abstract
Purpose: To study the power of an 80-second multifocal pupillographic objective perimetry (mfPOP) test tailored to the ETDRS grid to diagnose age-related macular degeneration (AMD) by Age-Related Eye Disease Study (AREDS) severity grade. Design: Evaluation of a diagnostic technology.Entities:
Keywords: AMD, age-related macular degeneration; AREDS, Age-Related Eye Disease Study; AUC, area under the receiver operating characteristic curve; BCVA, best-corrected visual acuity; ETDRS-grid OCT; OFA, Objective FIELD Analyzer; PD, pattern deviation; ROC, receiver operating characteristic; SD, standard deviation; TD, total deviation; abs(DelayDiff), delay difference; macular function; mfPOP, multifocal pupillographic objective perimetry; mfVEP, multifocal visual evoked potential; multifocal pupillography; objective perimetry; rapid perimetry
Year: 2022 PMID: 36249701 PMCID: PMC9559873 DOI: 10.1016/j.xops.2022.100143
Source DB: PubMed Journal: Ophthalmol Sci ISSN: 2666-9145
Figure 1The Objective FIELD Analyser (OFA) device. A, Illustration of a test participant seated and looking into the OFA objective lenses. B, Diagram showing the dichoptic stimulus arrangement inside the OFA. The pseudorandomly sequenced brief-onset stimuli are presented on 2 displays. Cold mirrors direct the eyes to the 2 displays. Infrared video cameras capture images of the 2 pupils. The record of the pupil diameters and the stimulus histories are used to extract the responses to each of the up to 44 stimulus test regions per eye. See Figure 2.
Figure 2Representations of the 3 Objective FIELD Analyser (OFA) stimulus types. All stimuli shown are those for the left eye (OS); stimuli for the right eye (OD) were mirror image. A, Contours of the 44 stimulus regions of the widefield P129 stimulus. The transiently presented (33 ms) stimuli were delivered pseudorandomly, with a mean per-region interval of 4 s. In practice, overlapping regions are never presented concurrently. B, Stimuli from the left and right halves of the 5 rings of P129 stimuli showing their relative intensities. C, Same as for (B), but for the 44-region P131 macular stimulus array. D, M18 stimulus array. E, Numbering scheme of the M18 stimuli as presented in visual space to OS. F, The projection of the M18 stimuli onto the retina showing the relationship to the ETDRS grid (cyan).
Percentage Area under the Receiver Operating Characteristic Curve for Best-Corrected Visual Acuity and OCT 9-Region ETDRS Grid Data
| Age-Related Eye Disease Study Grade | ||||
|---|---|---|---|---|
| 1 | 2 | 3 | 4 | |
| BCVA | 57.6 ± 6.76 | 42.6 ± 7.77 | 72.0 ± 7.20 | 83.9 ± 9.98 |
| Thickness | 44.2 ± 7.30 | 65.4 ± 8.05 | 54.6 ± 11.8 | 90.2 ± 7.07 |
| log volume | 44.4 ± 7.39 | 65.3 ± 7.90 | 49.0 ± 11.9 | 90.2 ± 6.32 |
BCVA = best-corrected visual acuity.
Percentage Area under the Receiver Operating Characteristic Curve and Hedge’s g Values for Objective FIELD Analyzer M18 Delays
| N Worst Value | ||||
|---|---|---|---|---|
| Percentage Area under the Receiver Operating Characteristic Curve ± Standard Error | Hedge’s g Value | |||
| 3 | 9 | 3 | 9 | |
| Delay | ||||
| AREDS Grade 1 | 83.8 ± 4.38 | 84.3 ± 4.21 | 1.24 | 1.23 |
| AREDS Grade 2 | 69.1 ± 7.51 | 72.8 ± 6.65 | 0.81 | 0.84 |
| AREDS Grade 3 | 85.8 ± 5.87 | 86.8 ± 5.33 | 1.79 | 1.83 |
| AREDS Grade 4 | 91.5 ± 4.56 | 92.9 ± 3.93 | 2.32 | 2.48 |
| abs(DelayDiff) | ||||
| AREDS Grade 1 | 84.7 ± 4.45 | 85.1 ± 4.44 | 1.31 | 1.32 |
| AREDS Grade 2 | 65.7 ± 7.34 | 69.8 ± 7.05 | 0.50 | 0.62 |
| AREDS Grade 3 | 81.4 ± 6.97 | 81.7 ± 7.31 | 1.35 | 1.37 |
| AREDS Grade 4 | 80.3 ± 7.63 | 79.3 ± 8.56 | 1.48 | 1.44 |
abs(DelayDiff) = delay difference; AREDS = Age-Related Eye Disease Study.
Percentage Area under the Receiver Operating Characteristic Curve for Pooled Objective FIELD Analyzer M18 Delays
| Delay | N Worst Value | |||
|---|---|---|---|---|
| Percentage Area under the Receiver Operating Characteristic Curve ± Standard Error | Hedge’s g Value | |||
| 3 | 9 | 3 | 9 | |
| AREDS Grade 1–2 | 78.1 ± 4.58 | 79.8 ± 4.39 | 1.24 | 1.23 |
| AREDS Grade 1–3 | 79.8 ± 4.12 | 81.4 ± 4.00 | 1.79 | 1.83 |
| AREDS Grade 1–4 | 81.3 ± 3.81 | 82.9 ± 3.71 | 2.32 | 2.48 |
AREDS = Age-Related Eye Disease Study.
Percentage Area under the Receiver Operating Characteristic Curve for P129 and P131 Delay Measures
| N Worst Value | ||||
|---|---|---|---|---|
| Percentage Area under the Receiver Operating Characteristic Curve ± Standard Error | Hedge’s g Value | |||
| 5 | 15 | 5 | 15 | |
| P129 abs(DelayDiff) | ||||
| AREDS Grade 1 | 67.6 ± 6.60 | 66.0 ± 6.64 | 0.70 | 0.58 |
| AREDS Grade 2 | ||||
| AREDS Grade 3 | 87.2 ± 3.59 | 86.9 ± 3.61 | 1.58 | 1.54 |
| AREDS Grade 4 | ||||
| P131 Delay | ||||
| AREDS Grade 1 | 73.9 ± 4.93 | 72.8 ± 5.13 | 0.97 | 0.95 |
| AREDS Grade 2 | ||||
| AREDS Grade 3 | 82.0 ± 5.94 | 82.9 ± 5.79 | 1.75 | 1.74 |
| AREDS Grade 4 | ||||
abs(DelayDiff) = delay difference; AREDS = Age-Related Eye Disease Study. Boldface indicates statistical significance.