| Literature DB >> 36221361 |
Jinlie Long1,2, Xiuhong Wang3, Jianbo Yuan1,2, Jianru Yang1,2, Jie Zhou1,2, Yuan Chen4, Enxi Hu1,2, Yuanzhong Zhou3, Xun Min1,2.
Abstract
The reference intervals of complete blood count (CBC) parameters were commonly based on healthy individuals aged 20 to 79 years. However, these values are not optimal for correct clinical diagnosis in older individuals (e.g., 80-89 years). Although the reference intervals for this age group have been reported in China, there is no population-based report in Guizhou province. A total of 481 healthy adults (238 males and 243 females) aged 80 to 89 years were recruited from Affiliated Hospital of Zunyi Medical University in Guizhou. The CBC parameters were detected by Sysmex XN-9000 automatic hematology analyzer. The reference intervals of the components were analyzed according to the guidelines of International Federation of Clinical Chemistry. This study reported the reference intervals of CBC parameters. There were significant differences were examined in some reference intervals between the different gender groups, especially for RBC-related parameters. Compared with national standards, the most of all conventional reference intervals for CBC parameters were decreased. The present study provided the local reference intervals of CBC parameters for individuals aged 80 to 89 years in Guizhou, China. Some of our results were sex-specific, and most of our results show lower values while comparing with commonly used reference intervals in China. Therefore, more attentions should be paid to these differences, and accurate reference intervals will facilitate clinical diagnosis and decision-making in these populations.Entities:
Mesh:
Year: 2022 PMID: 36221361 PMCID: PMC9542839 DOI: 10.1097/MD.0000000000030859
Source DB: PubMed Journal: Medicine (Baltimore) ISSN: 0025-7974 Impact factor: 1.817
Sex-specific medians, IQRs and reference intervals for red blood cell-related parameters obtained from healthy adults aged 80 to 89 years.
| Parameter | Sex | Median | IQR | Reference interval | Lower limit with 90% CI | Upper limit with 90% CI | National standard | |
|---|---|---|---|---|---|---|---|---|
| RBC, ×1012/L | M | 4.53 | 4.15–4.97 | <.001 | 3.32–5.73 | 3.26–3.39 | 5.66–5.79 | 4.3–5.8 |
| F | 4.26 | 3.99–4.52 | 3.40–5.10 | 3.35–3.45 | 5.05–5.14 | 3.8–5.1 | ||
| Hb, g/L | M | 144 | 130–154 | <.001 | 107–177 | 105–109 | 175–179 | 130–157 |
| F | 132 | 123–140 | 106–156 | 104–107 | 154–157 | 115–150 | ||
| Hct, L/L | M | 0.43 | 0.40–0.46 | <.001 | 0.33–0.53 | 0.32–0.33 | 0.52–0.53 | 0.40–0.50 |
| F | 0.40 | 0.37–0.42 | 0.33–0.47 | 0.32–0.33 | 0.47–0.47 | 0.35–0.45 | ||
| MCV, fL | M | 94.6 | 91.6–97.3 | .239 | 81.9–107.4 | 81.2–82.6 | 106.7–108.1 | 82–100 |
| F | 94.3 | 90.7–97.0 | 83.6–104.3 | 83.1–84.2 | 103.7–104.8 | |||
| MCH, pg | M | 31.4 | 30.5–32.8 | <.001 | 26.7–36.4 | 26.4–26.9 | 36.1–36.7 | 27–34 |
| F | 31.0 | 29.9–31.9 | 27.0–34.7 | 26.8–27.2 | 34.5–34.9 | |||
| MCHC, g/L | M | 332 | 327–339 | <.001 | 313–353 | 312–314 | 352–354 | 316–354 |
| F | 328 | 322–334 | 310–346 | 309–311 | 345–347 | |||
| RDW,% | M | 13.3 | 12.9–14.0 | .317 | 11.4–15.7 | 11.3–11.5 | 15.6–15.8 | - |
| F | 13.2 | 12.8–13.9 | 11.6–15.1 | 11.6–11.7 | 15.0–15.2 |
CI = confidence interval, F = female, Hb = hemoglobin, Hct = hematocrit, IQR = interquartile range, M = male, MCH = mean corpsular hemoglobin, MCHC = mean corpsular hemoglobin concentration, MCV = mean corpuscular volume, RBC = red blood cell, RDW= RBC distribution width.
P-value <.001 between our results and the national standard.
Standard reference intervals for blood cell analysis of the People’s Republic of China’s health industry (WS/T 405-2012).
Sex-specific medians, IQRs and reference intervals for white blood cell-related parameters obtained from healthy adults aged 80 to 89 years.
| Parameter | Sex | Median | IQR | Reference interval | Lower limit with 90% CI | Upper limit with 90% CI | National standard | |
|---|---|---|---|---|---|---|---|---|
| WBC, ×109/L | M | 6.14 | 5.09–7.48 | .346 | 2.99–9.83 | 2.80–3.17 | 9.65–10.02 | 3.5–9.5 |
| F | 6.15 | 5.10–7.00 | 3.20–9.22 | 3.04–3.36 | 9.05–9.38 | |||
| NEUT, ×109/L | M | 3.54 | 2.89–4.45 | .233 | 1.45–5.94 | 1.33–1.57 | 5.82–6.07 | 1.8–6.3 |
| F | 3.42 | 2.77–4.29 | 1.22–5.98 | 1.09–1.34 | 5.85–6.11 | |||
| NEUT, % | M | 58 | 53–64 | .825 | 40.9–74.3 | 40.0–41.8 | 73.4–75.2 | 40–75 |
| F | 58 | 51–64 | 38.4–76.5 | 37.3–39.4 | 75.4–77.5 | |||
| LYMPH, ×109/L | M | 1.80 | 1.40–2.27 | .437 | 0.34–3.57 | 0.26–0.43 | 3.49–3.66 | 1.1–3.2 |
| F | 1.90 | 1.47–2.38 | 0.66–3.24 | 0.59–0.73 | 3.18–3.31 | |||
| LYMPH, % | M | 30 | 25–36 | .072 | 13.9–47.1 | 13.0–14.8 | 46.2–48.0 | 20–50 |
| F | 31 | 26–38 | 14.6–49.0 | 13.7–15.6 | 48.1–49.9 | |||
| MONO, ×109/L | M | 0.52 | 0.43–0.63 | <.001 | 0.22–0.85 | 0.21–0.24 | 0.83–0.87 | 0.1–0.6 |
| F | 0.47 | 0.37–0.57 | 0.20–0.75 | 0.19–0.22 | 0.74–0.77 | |||
| MONO, % | M | 8 | 7–9 | <.001 | 4.6–12.3 | 4.4–4.8 | 12.1–12.5 | 3–10 |
| F | 7 | 6–9 | 3.9–11.7 | 3.7–4.1 | 11.5–11.9 | |||
| BASO, ×109/L | M | 0.05 | 0–0.07 | .820 | 0.00–0.11 | 0.00–0.00 | 0.11–0.11 | 0–0.06 |
| F | 0.05 | 0–0.07 | 0.00–0.11 | 0.00–0.00 | 0.10–0.11 | |||
| BASO, % | M | 1 | 0–1 | .911 | 0.00–1.63 | 0.00–0.00 | 1.58–1.69 | 0–1 |
| F | 1 | 0–1 | 0.00–1.62 | 0.00–0.00 | 1.57–1.67 | |||
| EO, ×109/L | M | 0.15 | 0.09–0.23 | .045 | 0.00–0.54 | 0.00–0.00 | 0.52–0.56 | 0.02–0.52 |
| F | 0.13 | 0.07–0. 20 | 0.00–0.42 | 0.00–0.00 | 0.41–0.44 | |||
| EO, % | M | 2 | 1–4 | .049 | 0.00–7.99 | 0.00–0.00 | 7.72–8.26 | 0.4–8.0 |
| F | 2 | 1–3 | 0.00–6.90 | 0.00–0.00 | 6.67–7.14 |
BASO = basophil, CI = confidence interval, EO = eosinophil, F = female, IQR = interquartile range, LYMPH = lymphocyte, M = male, MONO = monocyte, NEUT = neutrophil, WBC = white blood cell.
P-value <.001 between our results and the national standard.
Standard reference intervals for blood cell analysis of the People’s Republic of China’s health industry (WS/T 405-2012).
Sex-specific medians, IQRs and reference intervals for platelet -related parameters obtained from healthy adults aged 80–89 years.
| Parameter | Sex | Median | IQR | Reference interval | Lower limit with 90% CI | Upper limit with 90% CI | National standard | |
|---|---|---|---|---|---|---|---|---|
| PLT, ×109/L | M | 172 | 145–203 | <.001 | 84–267 | 79–89 | 262–272 | 125–350 |
| F | 196 | 160–239 | 88–316 | 81–94 | 309–322 | |||
| PDW,% | M | 14.0 | 12.1–16.1 | .666 | 8.4–20.4 | 8.0–8.7 | 20.0–20.7 | - |
| F | 13.7 | 12.1–15.8 | 8.7–19.7 | 8.4–9.0 | 19.4–20.0 | |||
| MPV, fL | M | 11.2 | 10.4–12.0 | .958 | 9.0–13.6 | 8.9–9.1 | 13.5–13.7 | - |
| F | 11.2 | 10.5–12.0 | 9.1–13.4 | 9.0–9.2 | 13.3–13.6 | |||
| P_LCR | M | 34.9 | 28.2–41.2 | .882 | 17.2–53.5 | 16.3–18.2 | 52.5–54.5 | - |
| F | 34.5 | 28.9–41.0 | 18.0–52.5 | 17.0–18.9 | 51.6–53.5 | |||
| PCT | M | 0.19 | 0.17–0.22 | <.001 | 0.10–0.29 | 0.10–0.11 | 0.28–0.29 | - |
| F | 0.22 | 0.18–0.26 | 0.11–0.34 | 0.11–0.12 | 0.33–0.34 |
CI = confidence interval, F = female, IQR = interquartile range, M = male, MPV = mean platelet volume, PCT = plateletcrit, PDW = PLT distribution width, P_LCR = PLT-larger cell ratio, PLT, platelet.
P-value <.001 between our results and the national standard.
Standard reference intervals for blood cell analysis of the People’s Republic of China’s health industry (WS/T 405-2012).